• Complies to ISO 11607-2 validation requirements for sealing.
• Critical process parameters to be controlled and monitored:
• Vacuum / gas level
• Seal temperature
• Seal time
• Seal pressure
• Automatic stop of cycle if one of the critical parameters has exceeded alarm limit.
• Critical process parameters can be calibrated.
• Seal data can be logged and exported to PC via USB stick.
• Equipped with accurate Audion Temperature Controller (ATC).
• Equipped with user friendly touchscreen interface.
• Access level controllable by login passwords for different user levels.
• Bi-active seal bars
• Seal width 8mm
• Gas flush
• Multicycle
• User administration function with personal passwords.
• Calibration
• Temperature meter for seal temperature verification (ATM)
• Number of sealbars: 2
• Machine dimensions (WxDxH): depends on model
• Chamber dimensions (WxDxH): depends on model
• Effective chamber dimensions: depends on model
• Effective sealbar length: depends on model
• Vacuum level: 25-700 hPa
• Gas level: 75 – 700 hPa
• Seal temperature: 100-200 °C
• Seal time: 0.8-10.0 sec
• Seal pressure: 1000-1800 hPa
• Power supply: 380-400V-13ph-50/60 Hz
• Air supply: 0.3 MPa (3 bar)
SPECIFICATIONS
Product: Audionvac floor model validatable
Seal length: up to 1090 mm
Seal width: 8 mm
Position of the sealing bar:
Depends on model
Effective chamber size (LxWxH): Depends on model
Type of machine
Validatable/calibratable/medical, Vacuum chamber
Model: Floor
Covering: Stainless Steel
Pump capacity: Depends on model
Type of film: PA/PE
Thickness of film: 2 x 150 micron
Type of package: Bag