constant climate chamber
KBF LQC 240 series BINDER

Characteristics

  • Type:

    constant

  • Length:

    785

  • Temperature:

    Max.: 70 °C (158 °F)

    Min.: 0 °C (32 °F)

  • Width:

    650

  • Depth:

    485

  • Capacity:

    247

Description

BINDER KBF LQC series employs innovative photostability function for maintaining constant climate within the chamber. Photostability is governed using two 3D spherical sensors, one responding to visible light and one to UV light. For achieving temperature, humidity, and light simulation, appropriate light dosage is set, and on reaching the set limit, the system switches off. The use of Photostability ensures long-term stability and makes away with the use of water.

Important characteristics

Temperature range: 0 °C to 70 °C
Temperature range with light: 10 °C to 60 °C
Humidity range: 10 % to 80 % RH
2 positionable illumination cassettes with ICH-compliant UV/Vis light source
Independent light-dose control of UV-A and visible light with spherical sensors
APT.line™ preheating chamber technology
Humidity regulation with capacitive humidity sensor and vapor humidification
Inner chamber made of stainless steel
Intuitive touchscreen controller with time-segment and real-time programming
Internal data logger, measured values can be read out in open format via USB
Unit self-test for comprehensive status analysis
Tight-sealing inner door made of safety glass (ESG)
Avoidance of glass corrosion by special TIMELESS coating
2 stainless steel racks
Access port with silicone plug, 30 mm, left
4 stable castors, two with brakes
Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
Computer interface: Ethernet
Door heating

{{requestButtons}}
Favorites