Measuring with specified VDI 2630 accuracy at all positions
Exclusively available in Waygate Technologies' Phoenix V|tome|x premium performance CT system family, True|position and Easy|calib technology expand the measurement positions with specified accuracy to all positions which can be verified with the innovative patented Ruby|plate calibration phantom.
3D Computed Tomography metrology starts here
Thanks to the patented Ruby|plate calibration phantom with its calibration uncertainty of less than 1 micrometer, even at a maximum probing length of 130 mm, not only a high measurement accuracy of up to SD ≤ (3.8 + L/100 mm) µm for both positions referring to VDI 2630-1.3 guideline can be specified, but also remarkable SD ≤ (5.5+ L/50 mm) µm for all other positions besides the VDI positions can be reached within the V|tome|x M premium CT scanner. Additionally, by applying the fully automated Easy|calib, the VDI specification SD ≤ (3.8 + L/100 mm) µm can be met at any position with just a few minutes of calibrating.
Due to highly automated routines, the new Ruby|plate calibration phantom and True|positions' compensation of remaining mechanical positioning uncertainties as well as correction of thermal effects by using temperature sensors, these proprietary innovations bring CT metrology workflows and precision to a new performance level.