Chlorine analyzer with pH, ORP and temperature measurement
AQUACHLORE
HEITO
Aquachlore is an instrument dedicated to the control of water. It enables a continuous measuring of pH, active chlorine and temperature.
Manufactured in a IP 65 waterproof wall mounting case, it is suitable for a variety of applications such as drinking water installations, hot water circuits and swimming pools thanks to its wide measuring range (0.01 mg/l- 5mg/l) of active chlorine.
Plugging in two dosing pumps enables both chlorine and pH regulation.
Active chlorine is the most disinfectant part of chlorine. The amperometric cell with three electrodes measures the electric current induced by active chlorine concentration : a reference potential is applied to the silver electrode which induces a chemical reaction, specific to HClO, between the working electrode covered with a thin diffusion membrane and the auxiliary electrode.
The sensitive end of chlorine/temperature probe is interchangeable. Diameter : 50 mm. Easy calibration on one point.
The Temperature probe is inside the chlorine chamber.
The pH measuring chamber is made up of a glass electrode and a solid electrolyte KCl gel reference electrode. No reactive chemical required. Easy calibration on two points.
Manufactured in a IP 65 waterproof wall mounting case, it is suitable for a variety of applications such as drinking water installations, hot water circuits and swimming pools thanks to its wide measuring range (0.01 mg/l- 5mg/l) of active chlorine.
Plugging in two dosing pumps enables both chlorine and pH regulation.
Active chlorine is the most disinfectant part of chlorine. The amperometric cell with three electrodes measures the electric current induced by active chlorine concentration : a reference potential is applied to the silver electrode which induces a chemical reaction, specific to HClO, between the working electrode covered with a thin diffusion membrane and the auxiliary electrode.
The sensitive end of chlorine/temperature probe is interchangeable. Diameter : 50 mm. Easy calibration on one point.
The Temperature probe is inside the chlorine chamber.
The pH measuring chamber is made up of a glass electrode and a solid electrolyte KCl gel reference electrode. No reactive chemical required. Easy calibration on two points.
-
zoom