INTERNORMEN Technology

Calibration software
CALSOFT INTERNORMEN Technology

  • calibration software CALSOFT INTERNORMEN Technology
Quick and cost-efficient inspection of the CCS 2 or for its secondary calibration.
Automatic comparison of marginal values for the following channels: > 4 µm, > 4,6 µm, > 6 µm, > 6,4 µm, > 10 µm, > 14 µm, > 21 µm, > 37 µm.
Safe and simple handling due to the support of the bottle sampling system BBS 2 and the solution for secondary calibration CALSUS 01.


Functionality:
A purified bottle sample of the calibration suspension CALSUS 01 (based on ISO 11171) will be freed from gas and introduced into the CCS 2 using the BBS 2.
The stated numbers of particles of certain sizes contained in the CALSUS 01 will be entered in the calibration software's user-menu. After the particle counting is completed, the calibration values will be recalculated and suggested for acceptance.
This automatic secundary calibration does not replace the annual Factory Calibration (which also includes additional maintenance and a new calibration certificate).