The InLight Manual reader provides readout for InLight System dosimeters. The InLight System measures radiation exposure with aluminum oxide detectors (Al 2O3:C) read out by optically stimulated luminescence (OSL) technology. The reader stimulates the detector with a light emitting diode (LED) array causing it to luminesce in proportion to the amount of radiation exposure. The luminescence is detected and measured by the reader's photo multiplier tube using a high sensitivity photon counting system. A dose calculation algorithm is then applied to the measurement to determine exposure results. This nondestructive type of readout allows for dose verification through reanalysis.

InLight menu-driven software resides on an external PC and provides control over the setup, analysis, and data recording enabling dosimeter read out and reader quality control. Designed for small volume facilities, the InLight Manual Reader enables you to maintain your own in-house accredited dosimetry program. For personnel, area/environmental, and emergency response monitoring, or any radiation assessment application.

Purchase or lease of reader includes operating, analysis and dose calculation algorithm software, and a laptop computer.