The Memmert blanket warmer IFbw is a Class I medical device.
Control technology
ControlCOCKPIT: SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT: the process time does not start until the set temperature is reached
Calibration: three freely selectable temperature values
adjustable parameters: temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
adjustable parameters: temperature (Celsius or Fahrenheit), programme time, time zones, summertime/wintertime
Ventilation
Fan: forced air circulation in working chamber, fixed value 100 %
Safety
Temperature control: adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Autodiagnostic system: for fault analysis