Aluminum oxide sensor dew-point transmitter AMT
4-20 mARS-4852-wire

Aluminum oxide sensor dew-point transmitter - AMT - Alpha Moisture Systems Ltd - 4-20 mA / RS-485 / 2-wire
Aluminum oxide sensor dew-point transmitter - AMT - Alpha Moisture Systems Ltd - 4-20 mA / RS-485 / 2-wire
Aluminum oxide sensor dew-point transmitter - AMT - Alpha Moisture Systems Ltd - 4-20 mA / RS-485 / 2-wire - image - 2
Aluminum oxide sensor dew-point transmitter - AMT - Alpha Moisture Systems Ltd - 4-20 mA / RS-485 / 2-wire - image - 3
Add to favorites
Compare this product

Characteristics

Technology
aluminum oxide sensor
Output
4-20 mA, RS-485, 2-wire, Modbus RTU
Protection level
IP66, NEMA 4X
Other characteristics
compact, stainless steel
Applications
for trace moisture measurements, for compressed air, for process applications
Measuring range

Max.: 20 °C
(68 °F)

Min.: -120 °C
(-184 °F)

Accuracy (temperature)

2 °C
(35.6 °F)

Description

Product overview
The AMT dewpoint transmitter is a compact, loop-powered instrument for continuous trace moisture measurement in process gases and compressed air. It uses ultra-high capacitance aluminium oxide sensing technology for high sensitivity, fast response and stable long-term performance. The transmitter integrates RISC microprocessor control, self-diagnostics and AutoCal span correction to simplify field calibration and verification.

Key features
  • Ultra-high capacitance aluminium oxide sensor for superior sensitivity, accuracy and repeatability
  • Factory-configurable outputs: °C/°F dewpoint, ppm(V), ppb(V), g/m³, lb/MMSCF
  • Rugged stainless steel enclosure with IP66 / NEMA 4X protection; pressure rated to 35,000 kPa (350 bar)
  • Compact 2-wire 4-20 mA loop-powered design for continuous online monitoring
  • Low maintenance design with easy installation and optional sensor holder/sample cell for simplified removal and calibration
  • Supplied with calibration certificate traceable to National and International Humidity Standards at NPL
  • Transmitter can be installed up to 1,000 metres from compatible displays (DS1200, DS4000, DS1500)

Modbus RTU features
  • Optional Modbus RTU over RS485 for digital connectivity with PLC and SCADA systems
  • Unique TAG IDs for straightforward device identification and asset management
  • Remote access to calibration, sensor and software parameters for proactive maintenance
  • Daisy-chain installation via RS485 bus supports multi-transmitter networks

Technical specifications
  • Model: AMT
  • Measurement range: dewpoint from -120 °C to +20 °C (-184 °F to +68 °F)
  • Accuracy: ±2 °C (±3.6 °F) dewpoint
  • Output: 2-wire 4-20 mA loop-powered (optionally Modbus RTU over RS485)
  • Sensing technology: ultra-high capacitance aluminium oxide sensor
  • Processor & features: RISC microprocessor, high resolution, self-diagnostics, AutoCal span correction
  • Construction: stainless steel enclosure
  • Ingress protection: IP66 / NEMA 4X
  • Maximum pressure rating: 35,000 kPa (350 bar)
  • Electrical connection: secure industrial connector
  • Optional accessories: sensor holder / sample cell for improved sampling and easy sensor removal
  • Calibration: supplied with certificate traceable to National and International Humidity Standards at NPL
  • Compatibility: compatible with DS1200, DS4000 and DS1500 displays; transmitter may be sited up to 1,000 metres from a display
  • Typical applications: continuous trace moisture monitoring in process systems and compressed air
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.