The exhaust ozone detector calculates the current ozone concentration value by measuring the change in light signal intensity before and after the absorption of UV light by ozone. We can offer OEM/ODM service.
UVOZ-3000 Exhaust Ozone Detector
The exhaust ozone detector calculates the current ozone concentration value by measuring the change in light signal intensity before and after the absorption of UV light by ozone. The equipment uses a dual-optical UV light source system with a lamp intelligent management system inside, which can make the UV lamp quickly reach the measurement state after power on. Using foreign separated light cell technology, the equipment has the characteristics of no leakage, high pressure resistance, high flow sampling gas impact resistance, easy cleaning, easy maintenance, simple operation and low cost of use.
It is mainly used for continuous detection of ozone gas in various industrial environments, and can also be used to detect ozone gas in operating pipelines, containers and other environments, and is widely used in pharmaceutical, chemical, municipal and wastewater treatment industries for ozone generator outlet concentration and ozone generator output measurement.
Ozone analyzer with touch screen operation interface, real-time display of ozone gas concentration, with pressure and temperature compensation and display, automatic zero time and manual zero mode can be set. High concentration (g/Nm3) range with flow input function, real-time display of ozone generator output. With two sets of relay signal output, users can freely choose the high alarm point and low alarm point signal output connection.