The Calibration Kit has been created to provide all the necessary components to enable customers to verify the accuracy of the readings on the Buckleys Contact and Proximity Probes so that ‘in-house’ verification certificates can be issued.
Providing the readings are correct, the silver/silver chloride screw-in half cell should be checked by using K-SAT Reference Cells in a standard salt water solution.
To confirm the readings, three cells should be used and the sea water corrosion test mixture should be made up to a 3% solution.
Kit comprises:
3 x screw-on K-SAT reference electrode cells
Sea water corrosion test mixture
3 x Adaptor leads
1 x zinc test block
Full instructions for ‘in-house’ verification
Technical specifications:
Electrical connection: screw fit
Output: +196mV @ 25°C vs SHE
Accuracy: ±10mV @ 25°C vs SHE
Temperature coeff: -0.64mV per/°C
Temperature range: 0 - 40 °C
Working life: up to 1 year
Dimensions: 155mm long x 25mm diam.
Packed weight (calibration kit): 1.05Kg
Dimensions (packed): 26cm x 14cm x 30cm