CYTOTOXIC CHEMOTHERAPY SAFETY CABINETS ACCORDING TO EN 61010-1 CEI 66-5 NORMS AND DPR 303/56; DPR 577/82; MARKed.
Entirely built in cold press-bent and painted FE P01 zinc plated steel sheet (Skin passat), thickness 10/10 mm, after degreasing treatment with a first
application of epoxy primer, 2 subsequent applications of thermosetting powders and following cabinet 4 levelling supports at its passage in thermal
gallery at 200 °C. “Monolith” type construction, with base.
Hollow core and reversible doors, built as above (with security glass into the GLASS VERSION), equipped with safety lock 4 points handles and lock.
110° opening to enable shelf extraction without tilting.
Safety warnings indicate the presence of dangerous and toxic products as well as the capacity and limit of product to be stored.
• 3 container shelves in steel as above with 10/10 mm thickness to contain liquids in case of possible breakage or spilling of containers.
Rack adjustment of shelf height.
• Electric fan marked.
Folded carbon steel-sheet epoxy coated housing. Galvanized steel sheet impellerforward curved blades.
Backward curving or radial blade’s fans must always be ducted to exhaust systems which pressure drop can restrict air flow o to attain absorbed current
values which are given on motor rating. IP 44, 300 mc/h, 60 dB noise, 2900 rpm.
• Fuses protection 5 x 20 1 A RAPID F.
• Fan capacity 300 mc/h.
• Power supply 220/230 V 50 Hz single-phase.
• Absorption 90 W.