N2O nitrous oxide sensor Gasboard-2522
tunable diode laser absorption spectroscopyTDLASfor automotive

N2O nitrous oxide sensor - Gasboard-2522 - Cubic Instruments (Wuhan) Ltd. - tunable diode laser absorption spectroscopy / TDLAS / for automotive
N2O nitrous oxide sensor - Gasboard-2522 - Cubic Instruments (Wuhan) Ltd. - tunable diode laser absorption spectroscopy / TDLAS / for automotive
Add to favorites
Compare this product

Characteristics

Gas type
N2O
Technology
tunable diode laser absorption spectroscopy, TDLAS
Applications
for automotive, for gas analyzers, for emission gas analysis
Output
RS-485
Other characteristics
OEM
Measuring range (ppm)

Min.: 0 unit

Max.: 1,000 unit

Measuring range (%)

Min.: 0.3 %

Max.: 2 %

Description

Product Overview

Gasboard-2522 is a high-precision N2O gas sensor based on Tunable Diode Laser Absorption Spectroscopy (TDLAS) and employing a quantum cascade laser (QCL) source. It is designed for highly selective and sensitive N2O measurement in automotive exhaust emission monitoring under the latest Euro 7 performance requirements. Gasboard-2522 integrates a high-quality optical probe with a proprietary gas cell to deliver high measurement accuracy, high sensitivity, strong resistance to cross-gas interference, and long-term stable operation. Featuring a modular design, Gasboard-2522 supports easy installation and system integration. Calibration and configuration are available via an RS485 communication interface.

Applications include automotive exhaust emission monitoring, environmental monitoring, and laboratory research.
Features
  • QCL-based narrow-band laser absorption spectroscopy with ultra-high sensitivity for N2O measurement, resistant to interference from other gases, humidity, and particulate matter.
  • Ultra-fast response time of ≤ 3 seconds for real-time monitoring.
  • Full-scale linear digital output for consistent, repeatable, and traceable measurement.
  • High-performance optical probe for stable operation and long service life.
  • Modular design for easy system integration and future upgrades.

Specifications

Principle: Tunable diode laser absorption spectroscopy (TDLAS)
Measurement: N2O
Range: 0~1000 ppm
Accuracy: ±0.3%FS or 2% reading
Zero Drift: Resolution: 0.1 ppm
Response Time T90: T10~T90 < 3 s
Constant Temperature: 60℃
Warm-up Time: Operating Temperature: -7 ~ 50℃
Operating Humidity: 0~98% RH (non-condensing)
Operating Voltage: 12 V ±5%
Operating Current: Communication: RS-485
Dimension: L×W×H: 372×155×81 mm (excluding air nozzle & foot pad)
Storage Condition: -40 ~ 85℃; 0~98% RH (non-condensing)