NDIR Anesthesia Gas Sensor MED-MZ001
Discription
Cubic MED-MZ001 is an anesthesia multi-gas sensor developed for simultaneous and continuous monitoring of CO2, N2O and anesthetic agents including Sevoflurane, Desflurane and Isoflurane in patient respiratory gases. Utilizing advanced NDIR (Non-Dispersive Infrared) technology, MED-MZ001 provides precise measurements with rapid response for real-time gas monitoring. It features strong resistance to cross-gas interference, ensuring accurate measurements in complex gas mixtures. Equipped with integrated pressure and temperature compensation algorithms, MED-MZ001 delivers reliable measurements across varying environmental conditions. Its compact design enables seamless integration into anesthesia machines. MED-MZ001 provides a high-performance sensing solution for anesthesia monitoring applications in operating rooms and critical care environments.
Features
•NDIR (Non-Dispersive Infrared) technology
•Simultaneous detection of Sevoflurane, Desflurane, Isoflurane, N2O, and CO2 concentrations
•Strong resistance to cross-gas interference
•High accuracy measurement
•Fast Response
•Long-term stability
•Compact and lightweight design
Basic performance
- Detecting principle: NDIR (non-dispersive infrared) technology
- Measurement gas 1): Sevoflurane(C4H3F7O), Desflurane(C3H2F6O),, Isoflurane(C3H2ClF5O),CO₂, N₂O
- Measurement range: Sevoflurane: 0 - 8 %Vol; Desflurane: 0 - 18 %Vol; Isoflurane: 0 - 5 %Vol; CO₂: 0 - 15 %Vol; N₂O: 0 - 100 %Vol"
- Accuracy: Sevoflurane, Desflurane, Isoflurane: ±(0.15% Vol + 5% of reading);
CO₂: ±(0.2% Vol + 2% of reading); N₂O: ±(2 %Vol+ 2% of reading) @ N₂O≤85%Vol
- Resolution: Sevoflurane, Desflurane, Isoflurane: 0.15% Vol; CO₂, N₂O: 0.1% Vol
- Sampling frequency: 100Hz
- Startup Time: ≤30s
- Response time 2): Sevoflurane, Desflurane, Isoflurane: T90< 600ms; CO₂, N₂O: T90< 250ms
- Sampling method: Suction Air pump
- Protection level: IP65
- Output: RS232
- Dimensions: 45.5mm*32.0mm*29.5mm
- Lifetime: ≥10years
- Shell material: Photosensitive resin
Environmental conditions
Working temperature: 10 ~40℃ (50 - 104℉)
Storage temperature: -25 ~ 60℃(-13 - 140℉)
Working humidity: 10-98%RH (non-condensing)
Storage humidity: 10-90%RH (non-condensing)
Working pressure: 660~1060mbar (with pressure compensation)
Storage pressure: 500~1060mbar
Electrical specifications
Power supply: 5V±5% DC
Average Working Current 3): 300mA (@25℃,5V)
Remark:
1) The sensor can simultaneously detect anesthetic gas (one of sevoflurane, desflurane, isoflurane) and CO₂, N₂O concentrations.
2) The response time is measured by passing 120mL/min of gas into the sensor at a background gas concentration of 60% of the full scale.
3) The current is measured at room temperature condition, and it may vary at high temperature or low temperature condition.