Product OverviewAn intrinsically safe ultrasonic thickness gauge designed for reliable thickness measurements in hazardous gas Zones (0, 1 & 2), coal mines and combustible dust Zones (21, 22). Rugged handheld instrument with large colour LCD, advanced data logging, A-scan and high temperature measurement capability, suitable for use on fuel depots, road and vessel tankers, mines, chemical plants, oil and gas, refineries, pipelines, hazardous storage tanks and similar hazardous environments.
Key Features- Certified to ATEX, IECEx and UKEX for use in hazardous gas Zones (0, 1, 2), coal mines and combustible dust Zones (21, 22)
- Intrinsic safety protection eliminating need for hot work permits
- Continuous measurement on hot assets in-service using a high temperature Ex probe
- Three measuring modes for through-coat, various materials and levels of corrosion (Single Echo, Echo-Echo, Multiple Echo)
- Deep Coat function to ignore thick coatings
- Advanced Data Logging (Linear, 2D/3D, Templates) with A-scan and Auto-Log
- Radial Points to investigate around a measurement spot/area of interest
- Manual Measurement Mode with configurable gates and gain
- Live A-scans and live B-scans for visual verification
- Four function keys for easy control and dynamic views
- User Access controls to protect records
- Measurement setup save/restore and Measurement Freeze with Ref/Min/Max limits
- Rugged solid-state electronic instrument in IP67 rated enclosure
- Bluetooth connection for transferring data to CygLink software
- Available variants: SC, TC, PLUS and PRO with upgradeable features
Available Variants / Options- SC Variant (Single Crystal use only)
- TC Variant (Twin Crystal use only)
- PLUS Variant (Single & Twin Crystal use)
- PRO Variant (Single & Twin Crystal use) with advanced data logging, A/B-scan, manual gates, user access and high temperature capability
Caractéristiques / Spécifications techniques- Measuring Modes: Single Echo with Twin Crystal Probes; Echo-Echo with Twin Crystal Probes; Multiple Echo with Single Crystal Probes
- Materials: Sound velocity range 1000 m/s to 9000 m/s (0.0390 in/us to 0.3543 in/us)
- Accuracy: ±0.1 mm (±0.004”) or 0.1% of thickness measurement, whichever is greatest
- Resolution: 0.1 mm, 0.05 mm or 0.01 mm depending on probe type
- Probe Options: Single crystal and twin crystal probes; twin element high temperature Ex probe available
- Measurement Range in Steel: 0.8 mm to 250 mm (0.031 in – 10 in) depending on probe/configuration, material and temperature
- Connector / Probe Sockets: Single Dual Coaxial Connector
- Power Supply: Rechargeable, removable Lithium-Ion battery pack
- Power Rating: 2 W
- Battery Life: 6-8 hours continuous measurement
- Display: 3.5" VGA sunlight readable colour display
- Size: 270 mm tall, 135 mm wide, 80 mm deep
- Weight: 1 kg (gauge with battery); 5 kg (complete gauge kit)
- Operating Temperature: -0°C to +50°C (32°F to 122°F)
- Storage Temperature: -10°C to +65°C (50°F to 149°F)
- Data Logging: Up to 10,000 measurements and A-scans per record; Linear, 2D/3D grids and templates supported; Auto-log feature; saves measurements and A-scans as records on internal memory
- Computer Software: CygLink for remote logging, viewing A-scans, survey and PDF report generation, graphic analysis and statistical calculations; Bluetooth connection for data transfer
- Certification: ATEX, IECEx and UKEX. Markings include I M1 Ex ia Ma; II 1G Ex ia IIC T4 Ga; II 2D Ex ib IIIB T100C Db; ambient range 0°C to +50°C; Certificate numbers highlighted by manufacturer
- Environmental Protection: IP67; Pollution degree 3
- Standards: BS EN ISO 16831
- Warranty: 3 years on gauge and 6 months on probe