The Rosemount 5081 Explosion-Proof Transmitter is a single channel loop powered device that can accept input from a variety of liquid analysis sensors including: pH/ORP sensors, conductivity sensors (contacting and toroidal) and amperometric sensors (dissolved oxygen, chlorine and ozone). The transmitter features a rugged, weatherproof, corrosion-resistant enclosure to withstand harsh environments.
Specifications
Operating Temperature-4 - 149 °F (-20 - 65 °C)
Power SupplyMinimum power supply voltage is 12 Vdc. Maximum power supply voltage is 42.4 Vdc (30 Vdc for intrinsically safe operation). Power supply voltage should cover the voltage drop on the cable plus the external load resistor required for HART® communications (250 Ω minimum).
Certifications/ApprovalsFor details, see specifications for the measurement of interest.
Enclosure RatingCast aluminum containing less than 6% magnesium, with epoxy polyester coating. NEMA 4X (IP65). Neoprene O-ring cover seals.
Weight10 lb/11 lb (4.5/5.0 kg)
Features
Robust explosion-proof design constructed with heavy duty cast aluminum
Faster installation using Quick Start programming
Intrinsically safe, non-incendive, and explosion-proof hazardous area location approvals
Visibility of process parameters by utilizing HART® or Profibus DP digital communications
Effortlessly connect with PLCs and DCSs by choosing the HART® or FOUNDATION™ Fieldbus communication options