These radiometers were developed for NIST, to be used with a broadband spectrometer to act as a spectral transfer standard when calibrating other detectors in the 0.25 to 15 μm range.
TEMPERATURE CONTROLLED POWER MEASUREMENT
Each head is composed of a low noise detector, thermistor, TE cooler and heatsink to compensate for any temperature change
THE ULTIMATE CHOICE IN MEASUREMENT STABILITY
Temperature control down to 0.05°C from 20 to 30°C gives a temperature coefficient <0.01 %, thus a voltage output stable to 0.01 %
2 SIZES AVAILABLE
• TP5-BL: 5 mm Ø pyroelectric sensor with organic black coating
• TP9-BL: 9 mm Ø pyroelectric sensor with organic black coating
TO BE USED WITH THE STEP-CONTROLLER
Plug your TP sensor into the STEP-Controller for power supply and temperature control. You can then use the analog output with a scope or lock-in