Temperature measuring system PA 40-K007
flowlaseroptical

Temperature measuring system - PA 40-K007 - Keller Pyrometer Systems - flow / laser / optical
Temperature measuring system - PA 40-K007 - Keller Pyrometer Systems - flow / laser / optical
Add to favorites
Compare this product

Characteristics

Measured physical value
temperature, flow
Technology
optical, camera-based, laser
Measured material
parts
Applications
thread
Configuration
linear

Description

Measuring system consisting of:
  • Pyrometer CellaTemp PA 40 AF 4
  • Mounting PA 20-066


Special features:
  • Two-colour pyrometer with focusable optics in a stainless steel housing
  • Broadband anti-reflective precision lens
  • 2 analogue outputs, 2 switching outputs, USB and RS 485
  • Sighting aid: through-the-lens sighting
  • Mounting fitting with cooling fitting and connection for an external fan for industrial use, with flexibly adjustable holding device
  • Switch cabinet with flow monitor and connection for water, air and mains voltage
  • External fan


Scope of delivery:
  • Pyrometer CellaTemp PA
  • Connection cable VK 02/A (5 m)
  • Video cable VK 02/F (5 m) for model with built-in camera
  • USB cable VK 11/D (1.8 m)
  • Software CellaView
  • Manual (short)


Technical data - Measurement parameters:
Measuring range: 500 - 3000 °C in several segments
Sensor: Double photo diode
Spectral range: 0.95 / 1.05 µm
Response time t98: Measuring range 500 - 1400 °C: ≤ 10 ms (T > 650 °C); Measuring range 650 - 1700 °C: ≤ 20 ms (T > 650 °C) and ≤ 10 ms (T > 750 °C); Measuring range 750 - 2400 °C: ≤ 10 ms (T > 950 °C); Measuring range 850 - 3000 °C: ≤ 10 ms (T > 1050 °C)
Measurement uncertainty: 1 % of temp. reading [°C]
Repeatability: 2 K
Temperature coefficient: 0.05 %/K
Resolution current output: 0.2 K + 0.03 % of selected range
Resolution interface: 0.1 K
Resolution display: 1 K

Optical variants (lenses and focusing ranges):
For measuring range 500 - 1400 °C: Objective F50 PZ 20.08: focusing 0.3 m - ∞
For measuring range 650 - 1700 °C, 750 - 2400 °C and 850 - 3000 °C: Standard objective lens PZ 20.01: focusing 0.4 m - ∞
Close-up lens PZ 20.03: focusing 0.2 m - 0.4 m
Wide-angle lens PZ 20.05: focusing 0.2 m - ∞
Telephoto lens I PZ 20.06: focusing 1.2 m - ∞
Telephoto lens II PA 20.06: focusing 0.6 m - ∞
Additionally for measuring range 850 - 3000 °C: Macro lens PA 40.01: focusing 0.086 m - 0.115 m

Electrical parameters:
Analog output: 0(4) - 20 mA linear, switchable, scalable
Resistance: ≤ 500 Ω
Analog input: 0 - 10 V
Switching output: 2 x Open Collector 24 V DC, ≤ 30 mA
Switching input: 2 x to 24 V
Interfaces: USB; RS 485
Power supply: 24 V DC +10 % / -20 %
Current consumption: ≤ 175 mA for 24 V DC

Camera:
Video output: Composite PAL, 1 Vpp, 75 Ω

Features:
Through-the-lens sighting; Laser pilot light; Camera option; 4-digit red LED display (digit-height 8 mm); 2 x LED status display; Control elements: 3 buttons

General parameters:
Connection: M12 thread 8-pin connection
Ambient temperature: 0 - 65 °C
Storage temperature: -20 - +80 °C
Permissible humidity: 95 % r.H. max. (non-condensing)
Dimension: Ø 65 mm x 220 mm (max. 277 mm)
Housing: Stainless steel V2A (1.4305)
Protection class: IP 65 acc. to DIN 40050
Weight: approx. 0.9 kg

Adjustable parameters (selection):
Measuring range; Measured variable and scale of analogue input and output signals; Smoothing filter; Memory operation mode and memory hold time; Emissivity; Transmittance; Compensation of reflected ambient radiation; Limits and operating mode of the alarm contacts; Parameters of the ATD function (optional); Linearization function; LED status display; Simulation of current and temperature; Temperature unit °C/°F; Measurement channel selection; Threshold level of contamination monitoring; For models with camera: TBC function and White balance

Functions (selection):
Overtemperature signal: if internal temp exceeds 80 °C the analogue output shows a value > 20.5 mA
Service function: Simulation of a measured value or a current signal
ATD (Automatic Temperature Detection) function: Automatic temperature detection at discontinuous processes

Measuring field (distance : diameter)
0.00 mm : 18.00 mm
16.00 mm : 17.39 mm
32.00 mm : 16.77 mm
48.00 mm : 16.16 mm
64.00 mm : 15.55 mm
80.00 mm : 14.93 mm
96.00 mm : 14.32 mm
112.00 mm : 13.71 mm
128.00 mm : 13.09 mm
144.00 mm : 12.48 mm
160.00 mm : 11.87 mm
176.00 mm : 11.25 mm
192.00 mm : 10.64 mm
208.00 mm : 10.03 mm
224.00 mm : 9.41 mm
240.00 mm : 8.80 mm
256.00 mm : 8.19 mm
272.00 mm : 7.57 mm
288.00 mm : 6.96 mm
304.00 mm : 6.35 mm
320.00 mm : 5.73 mm
336.00 mm : 5.12 mm
352.00 mm : 4.51 mm
368.00 mm : 3.89 mm
384.00 mm : 3.28 mm
400.00 mm : 2.67 mm

Software:
CellaView

Scope / Item information:
Item No.: 1093281

Catalogs

No catalogs are available for this product.

See all of Keller Pyrometer Systems‘s catalogs

Exhibitions

Meet this supplier at the following exhibition(s):

HärtereiPraxis 2026
HärtereiPraxis 2026

22-24 Jun 2026 Kamp-Lintfort im Wellings Parkhotel (Germany)

Shenzhen International Semiconductor & IC Exhibition
Shenzhen International Semiconductor & IC Exhibition

9-11 Sep 2026 Shenzhen (China)

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.