All-rounders for general use
Ideal for working with organic solvents and cold or diluted acids. Depending on the ceiling height of your laboratory, you can choose between tall or low designs.
Fume cupboard for low ceiling height with unique low installation height (2250 mm)
Above-average usable worktop depth
Underbench storage for tall utensils up to 720 mm
Metallic construction enables columns with hollow spaces at the front – for easy maintenance and protection against air leakage
Interior: coated steel, optional lining in Trespa®
Side or front service connections for easy tapping of water, gas, etc.
• Width : 1200
• Front height : 2300/2550
• Chamber height : 2300/2600
• Usable space width : 1100 /1400/1700/1900/2000
• Usable space height : 1190/1490
• Usable space depth : 800
• Work height : 900
• Minimal permissable volume flow rate* : 450/530/680/800
• Specified volume flow rate : 480/600/720/840
• Maximum permissible volume flow rate : 1500
Dimensions (mm) / Volume flow rate (m³/h)
Your work is what matters
When selecting a fume cupboard, the most important considerations are the activities you do and the substances you use.
The certification according to DIN EN 14175 part 2 and 3 allows EXPLORIS® fume cupboards for general use to be used universally in the laboratory. Below you will find all information about dimensions and volume flow.
Fume cupboard control TouchTronic
Easy operation of fume cupboards
As a control element for a wide range of functions, the new TouchTronic combines fume cupboard control and monitoring with high computing power.