OverviewCentral high-purity water system Titan, available in Nova and Classic families. Designed to supply ASTM Type II or Type I purified water with integrated 500 L stainless steel storage for laboratory, clinical and industrial applications.
Models & key features- Variants: Nova C300 / Nova C500 and Classic C300 / Classic C500
- Production rates: 300 L/h (C300) or 500 L/h (C500)
- Purification: RO + EDI with ≥ 99% rejection; produces ASTM Type II / Type I water
- Integrated pre-treatment module and 500 L stainless steel tank
- Dual-wavelength UV disinfection (254 nm & 185 nm) and microporous filtration to 0.2 µm
- Control and monitoring: TFT touchscreen, USB data logging and U-cloud remote monitoring (Wi‑Fi)
Main configurations- Raw water tank and raw water pump
- Sand/carbon filters and pre-treatment module; salt tank and resin softener
- Ultrafiltration membrane and high-pressure pump
- Reverse osmosis membrane followed by EDI module (main system)
- Water distribution pump, microporous pipeline filters and 0.2 µm inlet air filter
- Dual-wavelength UV lamp (254 nm & 185 nm) and 500 L stainless steel storage tank
- Pipeline connection: 3/4 inch
Technical specifications (summary)- Source water: Potable tap water
- Feed conductivity: < 2000 µS/cm
- Feed TOC: < 1 ppm
- Feed hardness*: < 450 ppm as CaCO3
- Operating pressure: 0.1–0.4 MPa (7–72 psi)
- Temperature range: 5–45 °C
- pH range: 4–10
- Resistivity at 25 °C**: > 8 MΩ·cm (typically 10–15 MΩ·cm)
- Conductivity at 25 °C**: 0.125 µS/cm (typically 0.1 µS/cm)
- Product TOC: < 30 ppb
- Particulates > 0.22 µm: < 1 pc/µL
- Bacteria: < 0.01 cfu/mL (< 10 cfu/L)
- RNase: < 1 pg/mL; DNase: < 5 pg/mL; Proteases: < 0.15 µg/mL
- RO rejection: ≥ 99%; EDI ion rejection: ≥ 99%
- Instantaneous demand flow: 500–1000 L/h***
- Electrical: 110/220 V ±10%, 50/60 Hz
- Dimensions & weights (examples): Main host 1750 × 1150 × 780 mm, net weight 380 kg; pre-treatment 1500 × 790 × 660 mm, weight 200 kg; 500 L tank 2150 × 700 × 700 mm, weight 57 kg
Applications and compliance- Suitable for research laboratories, clinical and diagnostic facilities, pharmaceutical support and industrial processes requiring ASTM Type I/II water
- Meets performance targets for resistivity, TOC, particulates and microbial limits when feed water meets specified quality
Installation & maintenance- Requires potable tap water feed; if hardness exceeds limits use pre-treatment with salt tank/resin softener*
- Recommended feed pressure 0.15–0.35 MPa for specified instantaneous flow
- Includes TFT control for local operation and U-cloud for remote monitoring; USB port for data logging
Options & services- Remote monitoring via U-cloud (Wi‑Fi module), additional storage tanks, spare parts and consumables
- Commissioning, on-site validation, user training and maintenance contracts available