- Heated gas sample dilution probe for sampling and dilution of process gas
- With a dilution ratio of 30:1 as standard, optionally in the range from 10:1 to 500:1
- Dilution gas pre-heater preheats the dilution gas to the sample probe temperature
- Dilution unit with critical orifice is located in the heated part of the sample probe
- With test gas connection at the sample probe
- Based on gas sample probe SP2000H
- Filter housing and mounting flange made of stainless steel can be heated up to max. 180 °C [356 °F]
- Ceramic filter element with a filter porosity of 2 µm
- Including weather protective cover
- Easy installation and low maintenance
- Low dead volume
- Power: 230 V
- Material of sample-contacting parts: stainless steel, quartz glass, Viton®, ceramic
- Optionally: add-on set A with 1 precision pressure controller and 2 vacuum pressure gauges; control panel S with 1 pressure controller and 2 pressure gauges, 1 flow meter and 2 shut-off valves
The electrically heated M&C dilution probe is used in such processes, where the measuring procedure or the handling of the process gas requires the dilution of the sample gas or the component(s) to be measured, e.g. for extremely toxic gases, for moisture or emission measurement in flue gases. Since the M&C dilution probe SP2000H/DIL... is based on the well-proven M&C gas sample probe SP2000H, a variety of applications requiring special filter techniques, materials, etc. can be handled without any problems with this dilution probe.