IntroductionThe GTR-V1 Gas Transmission Rate Tester is designed to measure gas transmission rates (O2, CO2, N2) of plastic films, composite films, high-barrier materials, sheets, metal foils, rubber, tires and permeable membranes using the differential pressure method. The unit is suitable for R&D, quality control and production monitoring.
Features- High-precision infrared sensor ensures accurate gas measurement
- Automatic humidity and temperature control stabilizes test environment
- Dedicated PC software centralizes test setup, execution and result visualization with real-time curves
- One-button automatic test operation for repeatable procedures
- Patented pipeline flushing design keeps carrier gas clean and improves testing efficiency
- Special sample sealing technology enhances chamber tightness and reduces leaks
- Optional computerized system available to meet USP 1207 requirements for data traceability and user rights management (pharmaceutical option)
Testing PrincipleThe pre-conditioned specimen is mounted in the diffusion cell to form a sealed barrier between two chambers. The lower-pressure chamber is evacuated, the entire cell is then evacuated, and gas is introduced into the higher-pressure side to create a constant pressure difference. Gas permeates through the specimen from high to low pressure; monitoring pressure changes in the lower chamber yields permeability and barrier properties.
Test standardASTM D1434, ISO 2556, ISO 15105-1
ConfigurationStandard configuration includes: main instrument, constant temperature controller, built-in computer, special sampler, fast quantitative filter paper, etc. Note: air source interface uses Φ6 mm tubing; gas source and distilled water supplied by user. Appearance shown is standard; customized configurations may vary.
Technical specifications- Model: GTR-V1
- Measurement range (film): 0.1~50,000 cc/㎡·24h·0.1MPa
- Sample quantity: 1
- Main sensor quantity: 1
- Test mode: single chamber independent
- Vacuum resolution: 0.01 Pa
- Vacuum degree: <10 Pa
- Temperature control range: 15℃~55℃
- Temperature control precision: ±0.05℃
- Temperature resolution: 0.01℃
- Test pressure: 0 MPa~0.2 MPa (regular)
- Air source pressure: 0.1 MPa~0.8 MPa
- Test gases: O2, N2, CO2 or other gas
- Interface of carrier gas: Φ6
- Vacuum tube diameter: 1/4"
- Calibration mode: calibration with standard film and external calibration of vacuum sensor
- Sample thickness: ≤3 mm
- Sample size: Φ97 mm
- Permeability area: 38.48 cm2
- Size (L×W×H): 765 mm × 440 mm × 470 mm
- Weight: 60 Kg
- Power supply: AC100~240V 50/60Hz