PW-CTS2-19-40A R & D two-box air-cooled thermal shock test chamber is launched for the materials research and development sector 19L small cold and thermal shock test chamber, self-research temperature balance technology, different from the traditional use of high-power refrigeration against high-power heating freezing balance technology.
Take the on-off refrigeration energy regulation technology – central controller according to the different temperature points to control the size of the cooling capacity to achieve temperature static balance, while allowing the operation of equipment is always in a relatively low-power state, energy saving and environmental protection.
1. High temperature tank impact range: +60 ℃ ~ +150 ℃
2. Low temperature tank impact range: 0℃ ~ -40℃
3.Temperature uniformity: ≤±2℃
4. Temperature fluctuation: ≤ ± 0.5℃.
5. High and low temperature shock constant temperature time: 5 min
6. Basket transfer time: within 5 seconds
(Temperature performance test is measured according to the relevant provisions of IEC60068-3 standard; the sensor is placed at the unit air outlet)
1. Studio size: W325 × H240 × D240 mm (internal box size can be customized)
2. External dimensions: approx. W750 × H1810 × D1120mm
3. Box structure: with independent product testing area, high temperature heat storage area, low temperature storage area
4. Inner box material: stainless steel plate (SUS 304#)
5. Outer box material: steel plate baking paint
6. Insulation material: high temperature tank: 24k glass wool
Low temperature tank: PU foam + glass wool
7. Heater: bare wire type electric heater
8. Air supply circulation system
a.TECO motor