Short description- Quantify optical quality of transparent materials such as plastic films.
- Measure haze and total transmission according to ASTM D1003 (CIE C).
- Rhopoint Novo-Haze TX provides single-measurement characterisation of transmission appearance.
What is transmission haze?Transmission haze is the scattering of light as it passes through a transparent material. It affects clarity, contrast and perceived quality of products viewed through films or sheets. Total transmission measures the fraction of incident light that passes through a sample, influenced by absorption and reflectance.
Transmission haze measurementHaze quantifies forward light scattering caused by internal particles, inclusions or fine surface texture. Haze testing is used to control optical properties of plastics, packaging films and coated substrates where reduced clarity can affect product presentation or functionality.
Instrument overviewThe Rhopoint Novo-Haze TX delivers fast, repeatable measurement of total transmission and haze for plastic films and other transparent materials. Designed to ASTM D1003 (CIE C), the instrument suits QA/QC and R&D environments and focuses on a cost-effective configuration dedicated to haze/transmission testing.
User interface and operationThe unit features an intuitive touch-sensitive display to minimise test time. All measurement functions are accessible from a single screen. Measurements can be triggered by touch or optional footswitch; batch testing provides basic statistics. Results can be printed or saved to USB for traceable records.
Included with the instrument- 1 x haze checking standard, NPL traceable
- Calibration certificate
- 24 V DC power supply with mains cable
- USB drive
Optional extras- Result printer
- Footswitch
- Custom sample holders
Applications- Packaging: clarity and consumer-visible quality of packed products
- Plastic films and flexible substrates
- Sheeted film and coated film applications
Technical specifications- Light source: White LED filtered for equivalence to CIE C luminosity function and human photopic response
- Measurement area: 14 mm Ø
- Measurement range: 0 - 100%
- Geometry: 0° / diffuse
- Sample thickness: Maximum 3 mm (standard instrument)
- Measurement standards: Conforms to ASTM D1003 (CIE C); verified to ISO 14782, ISO 13468-1, JIS K 7361
- Memory: 10,000 results via supplied USB stick; internal memory 99 results (statistical use)
- Power supply: 230 V / 50 Hz or 115 V / 60 Hz
- Operating conditions: Operating temperature 15 to 40 °C; storage temperature 0 to 50 °C
- Dimensions: 580 mm (H) x 200 mm (W) x 320 mm (D)
- Packed dimensions: 720 mm (H) x 440 mm (W) x 320 mm (D)
- Weight: 16 kg
- Packed weight: 21.7 kg
- Commodity code: 9027 5000
- Order code: A3000-001
- Haze performance: Repeatability ±0.05 units SD; Reproducibility ±0.4 units SD; Resolution 0.01%
- Transmission performance: Repeatability ±0.02 units SD; Reproducibility ±1.0 units SD; Resolution 0.01%