The 3010 MINIFID Portable heated FID VOC analyser is a small, self contained, and fully portable analyser. It is equally convenient for use on the laboratory bench or as a portable on-site measurement tool. This portable FID features the same detection system as the S4 Solar range of FIDs. It is supplied with a 5m heated sample line and built-in controller as standard. The bottle rack may be detached for use with standard, full sized gas cylinders.
Gases Sensed -
Methane (CH4), Non-methane Hydrocarbons (NMHC), Total Hydrocarbons (THC)
Technology -
Flame Ionisation Detection
Designed for portability
• Military style design
• Thousands sold globally
• Many still in operation after years of rigorous work
• Easy to use
• Warm-up typically 20 mins
• No microprocessor or software
• Just 5 simple functions
• 0-10v output or optional logger
Relevant Standards
• EN12619
• EN15267-3
• EN15267-4
• ISO25140
• Method 25
• Method 25a
• EURO VI/VII
• ARP1256D
• CPA 2015-C174
Applications
• Air separation
• Solvent/VOC abatement/scrubber efficiency
• Thermal oxidiser control/monitoring
• Process control
• Process emissions monitoring
• Flue gas emissions monitoring
• Combustion research
• Biomass combustion monitoring
• Carbon bed breakthrough detection
• Gas purity
• Locomotive/automotive/marine combustion engine exhaust emissions
Technical specification
SENSITIVITY: 0.1% of range
LINEARITY: 2% of measured value
DRIFT +/-0.2ppm or 2% range per week, whichever greater
NOISE +/-0.1ppm or 1% range, whichever greater
RESPONSE TIME: T90 in 2 seconds
POWER: 110V AC or 220V AC, 500 Watts
HEATED SAMPLE LINE TEMPERATURE: 190 o C