In-line Phase Switch for Turbidity or Colour
The PhaseGuard monitors phase transitions of turbidity or colour. Three models of the PhaseGuard cover the possible applications: phase switch for turbidity (model T), for colour (model C), for high turbidity (model HT). The PhaseGuard is factory adjusted with a universal measuring range based on absorption. For simple applications and system integration the instrument configuration and communication can be easily done using the serial interface in combination with the existing outputs. For a more comfortable installation and operation the optional control unit SICON with touch screen technology and colour display can be connected.
Sealless design
Extremely low maintenance
Recognition of the phase change by turbidity or colour
Wavelength turbidity LED 880 nm
Wavelength color LED 430 nm
Measuring principle Absorption
Nominal range 0 .. 100 % Absorption
Resolution 0.5 % Absorption
Installation In-line housing Varivent® or compatible
Enclosure Stainless steel 1.4301
Protection index IP66
Sample temperature -10 .. +100 °C
Pathlength 10 mm (models T & C)
5 mm (model HT)
Outputs 1 x 4 .. 20 mA
2 x Open-Collector Transistor
Pipe diameter
≥ DN 40 (modells T & C)
≥ DN 65 (model HT)
Supply voltage9 .. 30 VDC
Power input < 2 W (with Profibus DP: < 3 W)
Power input < 2 W (with Profibus DP: < 3 W)