An easy-to-use, high-performance calorimetry system. Using either Heat Flow Calorimetry (HFC) or Power Compensation Calorimetry (PCC), it can generate power and enthalpy graphs quickly and easily.
At a glance
The Atlas HD Reaction Calorimeter operates in either Heat Flow Calorimetry (HFC) or Power Compensation Calorimetry (PCC) mode. The system is designed for ease of use with the generation of power and enthalpy graphs made simple and quick.
In addition, the simple configuration of reaction temperatures, dosing quantities, and alarms, provides you with a safe and accurate reaction calorimetry system and offers great value for calorimetry chemists.
1. HFC and PCC measurements
The system offers two modes; Heat Flow Calorimetry (HFC) or Power Compensation Calorimetry (PCC), allowing you to choose which mode is suitable for your process. HFC is generally used for reactions where the heater may have an effect on the chemistry since the dip in heater is not used during the reaction. HFC is also preferred when large exotherms are expected. PCC is used where quicker results are required, since no pre- and post-calibration is required. All important reaction parameters can be monitored in real-time using the Atlas HD Calorimeter.
2. Powerful automation
The Atlas PC Reaction Calorimetry Software offers truly easy reaction calorimetry operation. Configuration of parameters such as reaction temperature, dosing quantities, expected power level (e.g. moderate endotherm; strong exotherm), and alarms (with subsequent actions) are defined in an easy to use wizard.