OverviewTA Instruments presents the Discovery X3 DSC, a differential scanning calorimeter with a multi-sample cell capable of measuring up to three samples simultaneously. Engineered for materials research, QC and method development, the X3 combines multi-sample throughput with advanced thermal measurement technologies and automation to streamline laboratory workflows.
Features and benefits- Fusion Cell™ with patented multi-sensor design for exceptional baseline flatness, sensitivity, resolution and reproducibility.
- Enhanced Tzero® heat flow technology delivering simultaneous temperature and enthalpy accuracy across all three samples.
- Three-sample calorimetry enabling replicate testing, statistical analysis and validation against a control sample.
- Modulated DSC® (MDSC®) to separate complex thermal events into reversing and non-reversing components.
- One-Touch-Away™ APP-style touch interface for instrument control and fast access to data on the instrument.
- Reliable linear autosampler with 54 programmable tray positions for unattended 24/7 operation and flexible experiment scheduling.
- Multiple refrigerated cooling options (sealed Refrigerated Cooling Systems) to reduce or eliminate liquid nitrogen use and extend sub-ambient operation.
- Tzero press and Tzero pans for reproducible sample encapsulation and optimal thermal contact.
- TRIOS software for instrument control, data analysis and reporting with one-click analysis, automated calibration routines and multi-instrument support.
- Manufacturer warranty covering cell and furnace to support instrument uptime.
Fusion Cell & Technology Highlights- Fusion Cell™ integrates proven Discovery and Q‑series features with a fixed X3 sensor for a stable, reproducible heat flow path and a uniblock silver furnace for improved temperature control.
- Tzero® technology measures sensor resistance and capacitance and accounts for pan contributions to provide flat baselines, high sensitivity and direct heat capacity measurement without baseline subtraction.
- Modulated DSC® uses a sinusoidal temperature modulation over a linear ramp to deconvolute total heat flow into reversing (heat capacity) and non-reversing (kinetic) components for clearer interpretation.
Automation, Autosampler & Accessories- Autosampler: rugged linear X-Y-Z design with integrated auto lid, laser positioning for calibration and pan verification, scheduled unattended calibrations/verifications, two quick-change trays and options to return or discard pans.
- Refrigerated Cooling Systems (RCS): sealed single-, two- or three-stage refrigeration to reach down to -40 °C, -90 °C or -120 °C for unattended experiments without liquid nitrogen.
- Liquid Nitrogen Pump (LN Pump) accessory for lowest operational temperatures (to -180 °C), high cooling rates and continuous operation with auto-fill Dewar capability.
- Sample preparation accessories: Tzero pans and lids, Tzero press, cutting and powder preparation kits for optimal sample contact and repeatability.
Software (TRIOS)- TRIOS software: integrated environment for instrument control, data acquisition, analysis and reporting; supports multi-instrument control, overlay/comparison across techniques and export to common formats.
- Advanced features: EasyMass integration for direct mass capture, JSON export for open-data workflows, batch processing, control charts and real-time/advanced analysis tools.
Technical specifications- Baseline Flatness (-50 to 300 °C): ≤ 10 μW
- Baseline Repeatability (-50 to 300 °C): < 15 μW
- Temperature Range: -180 °C to 550 °C
- Temperature Accuracy: ±0.025 °C
- Temperature Precision: ±0.005 °C
- Enthalpy Precision: ±0.08%
- Heating/Cooling Rate: 0.01 °C/min to 200 °C/min (depends on temperature and cooling accessory)
- Autosampler capacity: 54 programmable tray positions
- Included analyses: standard DSC (glass transition, peak integration, onset/endset, purity) and MDSC deconvolution into reversing and non-reversing heat flow.