Thermocouple wire for autoclave validation


Validation Thermocouple Wire and Looms (VTFP):
It is a regulatory requirement when validating autoclaves and lyophilizers that Type ‘T´ thermocouples of Class 1 material grade are used.

The VTFP validation thermocouple wire meets this requirement and is supplied either in pre-cut lengths from 1 metre upto 100 metre with a formed hot junction, the hot junction can be left bare or encapsulated in a Teflon tip. The cable can also be supplied on 100/300 metre long reels.

When considering the temperature mapping of large chambers and when multiple thermocouples are to be used, this material can be supplied as a cable loom. This approach eases the installation and general handling of multiple thermocouples and offers a more protective arrangement. All the conductors are identified with an ID number to ease location identity and connectivity.

When required the thermocouple cold junctions can be fitted with miniature or standard size thermocouple connectors.