The DB30 UV-Vis Spectrophotometer is a powerful double beam instrument designed for precise and accurate measurement of absorption and transmission across various wavelengths, making it essential for analytical applications where precision is paramount. Engineered for versatility, this user-friendly spectrophotometer excels in a wide range of measurements, from standard absorbance in cuvettes to solid samples and flow cell experiments. Its high-throughput optical design delivers exceptional performance, ensuring superior stray light control, baseline stability, and unmatched wavelength and photometric accuracy. The DB30 offers flexible operation, whether as a standalone device with its integrated touchscreen or through PC control using Visacle® software. Visacle provides a secure database for storing data and methods, and it can be upgraded to a CFR Part 11 compliant package for enhanced data protection and integrity.
Key Features- Variable bandpass options: 0.5, 1.0, 1.5, 2.0 or 4.0 nm over the range 190 nm – 1100 nm
- Fast scanning: up to 6,000 nm/min to aid sample analysis throughput
- Automatic accessory recognition when connected to the DB30
- Optional Visacle® software for easy PC control
- Touchscreen interface with intuitive menus and functions for ease of use for standalone control
- Multiple Data Storage Options: USB, SD card and internal data storage for convenient retrieval of methods and results
Software- Touchscreen for standalone operation controlling all modes: photometry, time scan, wavelength scan, and data analysis, including calibration curves
- Visacle® PC-controlled software for comprehensive control, validation, data analysis, and reporting
- Data securely stored in an organised database, ready for analysis and easy export in .txt, .csv, or PDF formats
- Visacle CFR Part 11 compliant package available with configurable user profiles, secure login, e-signatures, audit trail logging, and permission system
Measurement Examples- UV Absorption Spectra: Covers the UV range down to 190 nm with an absorbance range > 3, suitable for measuring samples with high UV absorption such as edible oils
Options and Upgrades- Standard Cell Holder: Two-position, holds traditional 10 mm path length cells; micro-volume cell holder option for 50 μl cells available
- Standard Cell Holder – Thermostatic: For incubation or temperature stabilisation from room temperature to +40°C, stability ±0.3°C
- Long Path Cell Holder: For low concentrations or absorbance, holds two rectangular cells with an optical path length of 10 mm to 100 mm
- Glass Filter Holder: For measuring transmittance/absorbance of glass samples or filters up to 55 mm x 100 mm, 5 mm thickness
- Film Holder: For thin-film samples up to 25 mm x 50 mm
- 6-Position Cell Holder: Mount up to six standard 10 mm path length cells with auto-changeover; temperature controlled version available
- Auto Sipper: For rapid and automatic measurement of multiple or large amounts of liquid sample without changing cells
- Micro Flow Cell: For continuous measurement of trace samples, 70 μl capacity, 10 mm path length with Teflon tubing
Customer Testimonial- “Our research lab is supported by a suite of Edinburgh Instruments products, enabling us to investigate a diverse range of materials, including perovskites, batteries, and 2D materials. The exceptional quality of their systems, combined with unmatched customer support, is why we continue to expand our lab with Edinburgh Instruments” – Dr Sajjad, Associate Professor, London South Bank University (RMS1000, FLS1000, DB30 user)
Technical Specifications- Optics: Czerny-Turner, Dual Beam Monochromator
- Wavelength Range: 190 nm - 1100 nm
- Spectral Bandwidth: 0.5 nm, 1 nm, 1.5 nm, 2 nm and 4 nm
- Stray Light: ≤0.10% T (220 nm NaI, 340 nm NaNO2)
- Wavelength Accuracy: ±0.1 nm D2 peak 656.1 nm
- Wavelength Repeatability: ±0.1 nm
- Photometric Range: Absorbance: -3.4 to +3.4, %T: 0 to 300, Concentration: 0.000 to 9.999
- Photometric Accuracy: ±0.002 Abs (0 - 0.5 Abs), ±0.004 Abs (0.5 - 1.0 Abs), ±0.008 Abs (1.0 - 2.0 Abs), ±0.3% T
- Photometric Reproducibility: ±0.0003 Abs (0-0.5 Abs), ±0.0005 Abs (0.5-1 Abs), ±0.001 Abs (1.0-2 Abs)
- Wavelength Scan Speed: 10, 100, 200, 400, 800, 1200, 2400, 3600, 6000 nm/min
- RMS Noise: ≤0.00003 Abs (0 Abs, 500nm)
- Baseline Stability: <0.0003 Abs/hr (500 nm, 2 hour lamp warm-up period)
- Baseline Flatness: ±0.0009 Abs (200 nm - 950 nm)
- Light Source: Tungsten-Halogen and Deuterium Lamps
- Light Source Switching: Automatic switching selectable for 325 nm - 370 nm range
- Detector: Silicon Photodiode
- Display: 7” Touchscreen
- Software: Touchscreen operation, PC-controlled software or CFR compliant PC-controlled software
- Dimensions: 500 mm (W) × 475 mm (D) × 250 mm (H)
- Net Weight: 20 kg (approx)
- Power Supply: 100 - 240V, 50/60 Hz, 150VA
- Ambient Temperature: 10°C - 35°C
- Output Device: USB flash drive, SD card
- Interface: USB PC interface