OverviewThe FAULHABER BHx A autoclavable drive solution is developed for medical handpieces and other sterilizable handheld devices. Its design supports repeated autoclave and disinfection cycles while maintaining low vibration, low surface temperature and reliable speed/torque performance at high rpm.
System Components- Brushless motors: 1645 BHS A, 1660 BHS A, 1660 BHT A
- Planetary gearhead: 16GPS A
- Encoders: IEM3-1024 A, IEM3-1024 LA
Key features- Designed to withstand more than 1,000 autoclave cycles
- Low surface temperature in operation
- Resistant to saturated-steam sterilization environments
- High speeds and high torque options for precision applications
Custom features- Mechanical and electrical interface adaptations
- Integration of thermal sensors
- Custom sealing solutions
Motors — summaryPlatform motors: 1645 BHS A, 1660 BHS A (high-speed variants) and 1660 BHT A (high-torque variant). Housing material: stainless steel.
Motors — performance tableMotor | Operation | Torque [mNm] | Power [W]
1645 BHS A | Continuous / Short-term | 8 / 12.3 | 57 / 99
1660 BHS A | Continuous / Short-term | 15.2 / 25.7 | 100 / 177
1660 BHT A | Continuous / Short-term | 18 / 30.3 | 88.6 / 161
Motor technical highlights- Number of pole pairs: 1
- Hall sensors: Digital
- Nominal voltage variants: 24 V, 36 V, 48 V
- Maximum speed: up to 100 000 min−1
- Housing material: Stainless steel
EncoderTechnology: Magnetic encoder. Output signals: Digital, 3 channels (A, B and Index). Resolution: 32 to 1,024 lines per revolution.
Gearhead: 16GPS ANumber of gear stages: 1 or 2. Reduction ratios (examples): 3.9:1, 4.9:1, 7.1:1, 15:1, 19:1, 24:1, 28:1, 35:1, 50:1. Length without motor: 18.9 mm (1 stage) / 26.1 mm (2 stages).
Gearhead performance tableGearhead | Operation | Max. input speed [rpm] | Max. torque [Nm] | Max. power [W]
16GPS A | Continuous / Short-term | 40,000 / 62,500 | 0.4 / 0.5 | 21 / 85
DownloadProduct brochure available (PDF).
Technical specifications- Product: FAULHABER BHx A autoclavable drive solution for medical handpieces
- Motors: 1645 BHS A, 1660 BHS A, 1660 BHT A
- Gearhead: 16GPS A (ratios up to 50:1)
- Encoders: IEM3-1024 A, IEM3-1024 LA
- Autoclave resistance: >1,000 cycles
- Surface temperature: low under typical operating conditions
- Environment: resistant to saturated steam sterilization
- Nominal voltages: 24 V / 36 V / 48 V
- Max. speed: up to 100,000 min−1
- Housing material: stainless steel
- Encoder resolution: up to 1,024 lines per revolution
- Customization: mechanical/electrical interfaces, thermal sensor integration, sealing solutions