HPW7000 HiPurity Water pH Measurement System provides accurate and stable pH measurement even in low conductivity (<10 mmho/cm) water samples. It is cost effective, easy to install and calibrate, it is considered an industry benchmark for accuracy.
The water being used to run expensive power generation turbines must be pure to ensure long life of the equipment. Good water chemistry control can minimize corrosion and prevent downtime due to maintenance. The HPW7000 HiPurity Water pH Measurement System is specially designed for this critical but difficult application
What Is It?
The HPW7000 is a specially designed electrode mounting assembly that provides accurate and stable pH measurement even in low conductivity (<10 mmho/cm) water samples. In addition to offering the lowest total cost of ownership, the device comes with a host of other benefits such as minimum calibration errors, improved temperature sensitivity, and small assembly that can be panel or surfaced mounted.
A flowing reference electrode design makes the mounting more compact in water sample panels where space is a constraint. Shielded electrode cables of up to 36 feet eliminate the need for mounting preamplifiers on the assembly and improve noise immunity. The HPW7000 can be connected to the UDA2182 and the DirectLine® Series for pH measurement.
How Does It Work?
The HPW7000 combines a specially designed stainless steel flow chamber with a measuring/reference electrode system and integral temperature sensor/solution ground.
Its specially designed flow chamber optimizes the sample flow volume and creates the proper flow and mixing patterns to make the most accurate and stable pH measurement.