LED pyrometer CellaTemp® PX 21
°CK180:1

LED pyrometer - CellaTemp® PX 21 - Keller Pyrometer Systems - °C / K / 180:1
LED pyrometer - CellaTemp® PX 21 - Keller Pyrometer Systems - °C / K / 180:1
LED pyrometer - CellaTemp® PX 21 - Keller Pyrometer Systems - °C / K / 180:1 - image - 2
LED pyrometer - CellaTemp® PX 21 - Keller Pyrometer Systems - °C / K / 180:1 - image - 3
Add to favorites
Compare this product

Characteristics

Display
LED
Units of measurement
°C, K
Optical resolution
100:1, 180:1, 50:1
Communication
analog output, IO-Link
Protection level
IP65
Configuration
stationary
Applications
industrial
Other characteristics
precision, with fiber optic, non-contact
Temperature

Max.: 2,000 °C
(3,632 °F)

Min.: 250 °C
(482 °F)

Spectral range

Max.: 1.7 µm

Min.: 1.1 µm

Response time

2 ms, 50 ms

Description

Product overview:
The CellaTemp PX 21 is a digital fibre-optic pyrometer for non-contact temperature measurement from 250 to 2000 °C. It features a separate optical measuring head connected by a fibre-optic cable, making the sensor immune to electromagnetic interference and allowing use of the electronics in ambient temperatures up to 250 °C without additional cooling. The separate measuring head variant is particularly suited for confined or hard-to-reach installation locations.

Special features:
  • Measuring range: 250–2000 °C
  • Modern IO-Link interface
  • Focusable, interchangeable lenses for exact adjustment of measuring distance
  • Broadband anti-reflective precision optics
  • High-contrast LED display readable from a distance
  • All parameters adjustable via front keyboard
  • Test current output for diagnostic checks
  • 1 analogue output and 2 switching outputs
  • Source freely configurable; ATD for automatic temperature detection
  • Measuring head and fibre cable heat resistant up to 250 °C


Versions / Optical variants:
Available versions: CellaTemp PX 21 AF 11, AF 21, AF 22.
Measuring range (all versions): 250–2000 °C.
Focus distances: AF 11: 0.2 m – ∞; AF 21: 0.12 m – ∞; AF 22: 0.033–0.045 m.
Distance ratio (D:S): AF 11: 180:1; AF 21: 100:1; AF 22: 50:1.
Measuring heads: AF 11 → PA 41.01 (M30); AF 21 → PA 41.05 (M16); AF 22 → PZ 41.18 (M16).

Measurement parameters:
  • Sensor: photodiode
  • Spectral range: 1.1–1.7 µm
  • Response time t98: ≤ 50 ms (T > 300 °C); ≤ 2 ms (T > 800 °C)
  • Measurement uncertainty: 0.75 % of reading, minimum 4 K
  • Repeatability: 2 K
  • Temperature coefficient: 0.25 K/K (T < 500 °C); 0.05 %/K (T > 500 °C)
  • Resolution current output: 0.2 K + 0.03 % of selected range
  • Resolution interface: 0.1 K; display resolution: 1 K


Electrical parameters:
  • Analogue output: 0(4)–20 mA, linear, switchable, scalable
  • Load resistance: ≤ 500 Ω
  • Switching outputs: 2 × open collector 24 V DC, ≤ 150 mA
  • Interface: IO-Link
  • Power supply: 24 V DC (+10 % / -20 %); current consumption ≤ 175 mA


Features (sighting / display / controls):
  • Sighting: laser pilot light
  • Display: 4-digit red LED (digit height 8 mm)
  • Status display: 4 × LED
  • Control: 3 front buttons; all parameters adjustable


General parameters:
  • Connection: M12, 5-pin
  • Ambient temperature: electronics 0–65 °C; measuring head heat resistant up to 250 °C
  • Storage temperature: -20 to +80 °C
  • Humidity: max. 95 % r.H. (non-condensing)
  • Dimensions (electronics): Ø 65 mm × 187 mm; housing stainless steel V2A (1.4305)
  • Protection class: IP65 (DIN 40050)
  • Weight: approx. 0.9 kg


Adjustable parameters:
  • Analog output scaling (0/4–20 mA), smoothing filter
  • Memory operation mode and hold time
  • Emissivity and transmittance
  • Alarm contact limits and mode
  • ATD function parameters; simulation of current/temperature; °C/°F


Functions:
  • Overtemperature signal (internal temp > 75 °C)
  • Service function: simulation of measured value or current signal


Scope of delivery:
  • Pyrometer CellaTemp PX 21 (electronic unit)
  • Connection cable VK 02/L AF 1 (5 m)
  • Measuring head (version-dependent)
  • Short manual
  • * Fibre optic cable must be ordered separately


Software:
SW 50 (configuration and data readout)

Versions / Measuring field (example AF 22 at 200 mm):
Measuring distance 0.00 mm → measuring field Ø 21.50 mm; 8.00 mm → Ø 20.68 mm; 16.00 mm → Ø 19.87 mm; … up to 200.00 mm → Ø 1.11 mm (see technical chart).

Technical specifications (summary):
  • Model: CellaTemp PX 21 (AF 11 / AF 21 / AF 22)
  • Measuring range: 250–2000 °C; spectral range 1.1–1.7 µm
  • Response time: ≤ 50 ms (T > 300 °C); ≤ 2 ms (T > 800 °C)
  • Measurement uncertainty: 0.75 % of reading (min. 4 K); repeatability 2 K
  • Analogue output: 0(4)–20 mA; switching outputs: 2 × OC 24 V DC
  • Interface: IO-Link; power: 24 V DC (+10 % / -20 %)
  • Housing: stainless steel V2A; protection class IP65; display: 4-digit red LED
  • Measuring heads: PA 41.01, PA 41.05, PZ 41.18
  • Scope of delivery: pyrometer, connection cable (5 m), measuring head, short manual
  • Item No.: 1097777

Catalogs

No catalogs are available for this product.

See all of Keller Pyrometer Systems‘s catalogs

Exhibitions

Meet this supplier at the following exhibition(s):

HärtereiPraxis 2026
HärtereiPraxis 2026

22-24 Jun 2026 Kamp-Lintfort im Wellings Parkhotel (Germany)

Shenzhen International Semiconductor &amp; IC Exhibition
Shenzhen International Semiconductor &amp; IC Exhibition

9-11 Sep 2026 Shenzhen (China)

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.