DescriptionCO₂ incubators must ensure stable, reproducible growth conditions for cell cultures while minimising contamination risk. The MicroSENS 100-VS is an NDIR CO₂ sensor designed for direct installation inside the incubation chamber. It provides fast, precise CO₂ and temperature feedback for reliable atmosphere control under real incubator conditions and supports advanced control strategies.
Key facts- Product type: NDIR gas sensor for CO₂ measurement in incubators, sensor head vapour-sterilizable at 100 °C
- Target application: Direct installation inside the incubation chamber to measure the actual atmosphere experienced by the cells
- Measuring range: 0 … 20 Vol.-% CO₂ (optimized for accurate monitoring up to 5 Vol.-%)
- Measurement principle: IR dual-beam technology (NDIR)
- Compensation: Temperature, pressure and humidity compensation
- Performance: Accuracy ± (2% of reading + 0.2 Vol.-%), repeatability < 0.11% FS, response time t90 < 25 s (with filter), digital resolution 0.001 Vol.-%
- Long-term stability: ≤ ± 0.2 Vol.-% at 5 Vol.-% / year
Characteristics / technical specifications- Maximum measuring range: 20 Vol.-%
- Measuring gas: carbon dioxide
- Gas supply: Diffusion
- Warm-up time at startup: < 15 s
- Heating time for full specifications: < 5 min
- IP protection class (default value): Sensor head IP54, area near connector plug IP30
- Repeatability: < 0.11 % FS
- Accuracy: ± (2 % of reading + 0.2 Vol.-%)
- Response time with filter: < 25 s
- Digital resolution: 0.001 Vol.-%
- Temperature influence: 0.125 % FS / 10 K
- Pressure influence: 0.004 % FS / 10 hPa
- Long-term stability: ≤ ± 0.2 Vol.-% at 5 Vol.-% / year
- Operating voltage (analog connection): 15 - 24 V DC
- Operating voltage (digital connection): 12 - 24 V DC
- Nominal power consumption: < 2 W
- Digital output signal: RS485
- Analogue output signal: Current 4 - 20 mA
- Minimum operating temperature: 0 °C
- Maximum operating temperature: 60 °C
- Minimum storage temperature: -20 °C
- Maximum storage temperature: 85 °C
- Maximum sterilization temperature: 100 °C (sensor head)
- Environmental humidity: 0 ... 95 % relative humidity (non condensing)
- Minimum air pressure: 500 hPa
- Maximum air pressure: 1,200 hPa