NDIR CO2 sensor MicroSENS 100-VS
infrared4-20 mARS-485

NDIR CO2 sensor - MicroSENS 100-VS - Micro-Hybrid Electronic GmbH - infrared / 4-20 mA / RS-485
NDIR CO2 sensor - MicroSENS 100-VS - Micro-Hybrid Electronic GmbH - infrared / 4-20 mA / RS-485
Add to favorites
Compare this product

Characteristics

Type
NDIR, infrared
Output
4-20 mA, RS-485
Other characteristics
temperature-compensated, pressure-compensated, with temperature measurement
Measuring range (%)

Max.: 20 %

Min.: 0 %

Description

Description
CO₂ incubators must ensure stable, reproducible growth conditions for cell cultures while minimising contamination risk. The MicroSENS 100-VS is an NDIR CO₂ sensor designed for direct installation inside the incubation chamber. It provides fast, precise CO₂ and temperature feedback for reliable atmosphere control under real incubator conditions and supports advanced control strategies.

Key facts
  • Product type: NDIR gas sensor for CO₂ measurement in incubators, sensor head vapour-sterilizable at 100 °C
  • Target application: Direct installation inside the incubation chamber to measure the actual atmosphere experienced by the cells
  • Measuring range: 0 … 20 Vol.-% CO₂ (optimized for accurate monitoring up to 5 Vol.-%)
  • Measurement principle: IR dual-beam technology (NDIR)
  • Compensation: Temperature, pressure and humidity compensation
  • Performance: Accuracy ± (2% of reading + 0.2 Vol.-%), repeatability < 0.11% FS, response time t90 < 25 s (with filter), digital resolution 0.001 Vol.-%
  • Long-term stability: ≤ ± 0.2 Vol.-% at 5 Vol.-% / year


Characteristics / technical specifications
  • Maximum measuring range: 20 Vol.-%
  • Measuring gas: carbon dioxide
  • Gas supply: Diffusion
  • Warm-up time at startup: < 15 s
  • Heating time for full specifications: < 5 min
  • IP protection class (default value): Sensor head IP54, area near connector plug IP30
  • Repeatability: < 0.11 % FS
  • Accuracy: ± (2 % of reading + 0.2 Vol.-%)
  • Response time with filter: < 25 s
  • Digital resolution: 0.001 Vol.-%
  • Temperature influence: 0.125 % FS / 10 K
  • Pressure influence: 0.004 % FS / 10 hPa
  • Long-term stability: ≤ ± 0.2 Vol.-% at 5 Vol.-% / year
  • Operating voltage (analog connection): 15 - 24 V DC
  • Operating voltage (digital connection): 12 - 24 V DC
  • Nominal power consumption: < 2 W
  • Digital output signal: RS485
  • Analogue output signal: Current 4 - 20 mA
  • Minimum operating temperature: 0 °C
  • Maximum operating temperature: 60 °C
  • Minimum storage temperature: -20 °C
  • Maximum storage temperature: 85 °C
  • Maximum sterilization temperature: 100 °C (sensor head)
  • Environmental humidity: 0 ... 95 % relative humidity (non condensing)
  • Minimum air pressure: 500 hPa
  • Maximum air pressure: 1,200 hPa

Catalogs

No catalogs are available for this product.

See all of Micro-Hybrid Electronic GmbH‘s catalogs
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.