DescriptionThe MicroSENS 180-HS-PFAS-FREE 0–20% CO₂ gas sensor is an NDIR diffusion sensor designed for CO₂ monitoring and control in cell incubators. The sensor head is placed inside the incubation chamber to measure CO₂ concentration directly in the cell culture environment using infrared absorption for reliable and precise determination of CO₂ concentration.
Key features- NDIR CO₂ sensor (IR dual-beam technology)
- PFAS-free design (no intentionally added PFAS in wetted/gas-exposed sensor head materials, including filter concept)
- Temperature and pressure compensation
- Humidity compensation
- Heat sterilizable at typ. 180 °C (short-term up to max. 190 °C)
- Long service life; designed for high sterilization cycle counts
- Digital (RS232) and analogue (4–20 mA) interfaces
Performance and integrationThe sensor uses a dual-beam NDIR method with low drift and low susceptibility to interference, delivering high reliability and long lifetime. Heat sterilization of the sensor head (typ. 180 °C, short-term up to 190 °C) reduces cross-contamination risk and simplifies decontamination. Temperature, pressure and humidity compensation ensure stable readings in humid incubator atmospheres. A 15-point calibration matrix (five concentrations at three temperatures) improves linearity over the full 0–20% range. Digital and analogue outputs and a feed-through mounting concept facilitate OEM integration with incubator control electronics.
Technical specifications- Minimum measuring range: 0 Vol.-%
- Maximum measuring range: 20 Vol.-%
- Gas supply: Diffusion
- Warm-up time at startup: < 10 s
- Heating time for full specifications: < 5 min
- IP protection class (default): IP 44
- Repeatability: < 0.005 % FS
- Accuracy: ± (2 % of reading + 0.2 Vol.-%)
- Response time with filter: < 100 s
- Digital resolution: 0.001 Vol.-%
- Temperature influence: 0.155 % FS / 10 K
- Pressure influence: 0.004 % FS / 10 hPa
- Long-term stability: ≤ ± 0.2 Vol.-% at 5 Vol.-% / year
- Operating voltage (analog): 15 - 24 V DC
- Operating voltage (digital): 12 - 24 V DC
- Nominal power consumption: < 2 W
- Digital output signal: RS232
- Analogue output signal: Current 4 - 20 mA
- Minimum operating temperature: 0 °C
- Maximum operating temperature: 60 °C
- Minimum storage temperature: -20 °C
- Maximum storage temperature: 85 °C
- Maximum sterilization temperature: 190 °C
- Environmental humidity: 0 ... 95 % relative humidity (non condensing)
- Minimum air pressure: 500 hPa
- Maximum air pressure: 1,200 hPa
- Measuring gas: carbon dioxide