Description
Product introduction
Mask respiratory resistance tester measures inspiratory and expiratory resistance of respirators and face masks under prescribed conditions. It is suitable for inspection agencies, manufacturers and quality control of ordinary masks, dust masks and medical masks.
Applications
Measurement of inspiratory and expiratory resistance for respirators and face masks.
Use in national labor protection equipment inspection, manufacturer production control and testing laboratories.
Technical features
Includes adjustable air source, standard head form, separate inhalation and exhalation tubing systems and respiratory resistance measurement system.
Thermal mass flowmeter with high sensitivity and very low starting flow; range 0–100 L/min; accuracy ±2%.
Multiple sensors on a single flow sensor chip to improve dynamic range; high zero stability, full-range stability, low power consumption and fast response.
Micropressure sensor range -1000 Pa to 1000 Pa; accuracy 1 Pa; two-wire design with explosion-proof housing for safety and reliability.
Special gas-path switching element for quick conversion between exhalation and inhalation lines.
Air pump capacity 40–100 L/min; constant ventilation mode (85±1) L/min.
Data recorder: 128×64 graphic dot-matrix LCD (Chinese interface); data logger memory 4M.
Power supply: AC 220 V, 50 Hz.
Applicable standards
GB 2626-2019 Respiratory protective equipment (inspiratory and expiratory resistance clauses)
GB/T 32610-2016 Technical specification for daily protective masks (inspiratory and expiratory resistance clauses)
GB 19083-2010 Technical requirements for medical protective masks
GB 2890-2009 Respiratory protection self-priming filter gas mask (clause 6.9)
Technical specifications
Model: MU-K1002
Flowmeter range: 0–100 L/min; accuracy ±2%
Flow sensor type: thermal mass flowmeter
Micropressure gauge range: -1000 Pa to 1000 Pa; accuracy 1 Pa
Air pump capacity: 40–100 L/min
Constant ventilation: (85±1) L/min
Data recorder: 128×64 graphic dot-matrix LCD (Chinese interface)
Data logger memory: 4M
Power supply: AC 220 V, 50 Hz