Hydrogen analyzer H₂-Trans
monitoringindustrialconcentration

Hydrogen analyzer - H₂-Trans - Pro-Chem-Analytik GmbH & Co. KG - monitoring / industrial / concentration
Hydrogen analyzer - H₂-Trans - Pro-Chem-Analytik GmbH & Co. KG - monitoring / industrial / concentration
Add to favorites
Compare this product

Characteristics

Measured entity
hydrogen
Application domain
monitoring, industrial
Measured value
concentration, trace
Configuration
compact
Technology
electrochemical
Protection level
IP65
Other characteristics
calibration, cost-effective, high-speed, ATEX

Description

Description
The H₂-Trans is an industrial hydrogen analyzer designed for accurate measurement at trace (ppm) and percent levels. The instrument offers a simple display and one-point calibration for fast commissioning. Optional ATEX certification: Ex ia IIC T4.

Overview
  • Measured parameters: Hydrogen (% and ppm)
  • Detection method: electrochemical fuel cell
  • Measuring ranges: user-selectable; 0–10,000 ppm and 0–100 % (user ranges between 100 ppm and 10 %)
  • Accuracy: ±2 % FSD
  • Response time T90: up to 45 s (0–1 %)
  • Output: 4–20 mA, 2-wire
  • Supply voltage: 10–35 V


Detailed description
The unit supports one user-defined measuring range within the available span. It is built for continuous industrial use with a compact enclosure suited to on-site monitoring and integration into process control systems.

Sensor
The H₂-Trans uses a dedicated fuel cell sensor optimised for hydrogen detection. The sensor provides the sensitivity and lifetime expected for industrial monitoring tasks.

Calibration
Trace-range calibration must be performed with a calibration gas at the chosen concentration inside the measuring range. For percent-range measurements, calibration can be performed using ambient air.

Features
  • Compact footprint
  • Cost-effective solution for continuous monitoring
  • Long-life sensor
  • IP65 enclosure for industrial environments
  • Reverse-voltage protection and temperature compensation


Flow-through measuring cell in Stainless Steel
  • Modular stainless steel measuring cell
  • Field serviceable: replace the defective part without replacing the entire cell


Standard Features
  • Interchangeable sensor options
  • Five selectable measuring ranges
  • 316 SS stainless steel cell block


Application Areas
  • Semiconductor industry
  • Gas manufacturers
  • Metallurgical industries
  • Chemical industry
  • Heat treating and bright annealing


Technical specifications
  • Parameters: Hydrogen % / Hydrogen ppm
  • Method: electrochemical (fuel cell)
  • Range: 0–10,000 ppm; 0–100 %; user ranges between 100 ppm and 10 %
  • Resolution: 0.1 ppm
  • Flow rate: approx. 1 L/min
  • Pressure: 0.1–1 bar
  • Calibration: calibration gas for trace range; air possible for percent range
  • Response time T90: 0–1 % = 45 s
  • Gas temperature: 0–50 °C
  • Output: 4–20 mA, 2-wire
  • Supply voltage: 10–35 V
  • Display: LCD 6 digits with bargraph
  • Protection class: IP65
  • Connection: 6 mm
  • Dimensions: 120 x 160 x 65 mm
  • Weight: 2 kg
  • ATEX: optional Ex ia IIC T4
  • SIL: Yes
  • Accuracy: ±2 % FSD