DescriptionThe H₂-Trans is an industrial hydrogen analyzer designed for accurate measurement at trace (ppm) and percent levels. The instrument offers a simple display and one-point calibration for fast commissioning. Optional ATEX certification: Ex ia IIC T4.
Overview- Measured parameters: Hydrogen (% and ppm)
- Detection method: electrochemical fuel cell
- Measuring ranges: user-selectable; 0–10,000 ppm and 0–100 % (user ranges between 100 ppm and 10 %)
- Accuracy: ±2 % FSD
- Response time T90: up to 45 s (0–1 %)
- Output: 4–20 mA, 2-wire
- Supply voltage: 10–35 V
Detailed descriptionThe unit supports one user-defined measuring range within the available span. It is built for continuous industrial use with a compact enclosure suited to on-site monitoring and integration into process control systems.
SensorThe H₂-Trans uses a dedicated fuel cell sensor optimised for hydrogen detection. The sensor provides the sensitivity and lifetime expected for industrial monitoring tasks.
CalibrationTrace-range calibration must be performed with a calibration gas at the chosen concentration inside the measuring range. For percent-range measurements, calibration can be performed using ambient air.
Features- Compact footprint
- Cost-effective solution for continuous monitoring
- Long-life sensor
- IP65 enclosure for industrial environments
- Reverse-voltage protection and temperature compensation
Flow-through measuring cell in Stainless Steel- Modular stainless steel measuring cell
- Field serviceable: replace the defective part without replacing the entire cell
Standard Features- Interchangeable sensor options
- Five selectable measuring ranges
- 316 SS stainless steel cell block
Application Areas- Semiconductor industry
- Gas manufacturers
- Metallurgical industries
- Chemical industry
- Heat treating and bright annealing
Technical specifications- Parameters: Hydrogen % / Hydrogen ppm
- Method: electrochemical (fuel cell)
- Range: 0–10,000 ppm; 0–100 %; user ranges between 100 ppm and 10 %
- Resolution: 0.1 ppm
- Flow rate: approx. 1 L/min
- Pressure: 0.1–1 bar
- Calibration: calibration gas for trace range; air possible for percent range
- Response time T90: 0–1 % = 45 s
- Gas temperature: 0–50 °C
- Output: 4–20 mA, 2-wire
- Supply voltage: 10–35 V
- Display: LCD 6 digits with bargraph
- Protection class: IP65
- Connection: 6 mm
- Dimensions: 120 x 160 x 65 mm
- Weight: 2 kg
- ATEX: optional Ex ia IIC T4
- SIL: Yes
- Accuracy: ±2 % FSD