- Perform AC and DC dielectric strength tests
- Up to 500VA (100mA) power output
- Tests to many international safety standards
The Sentinel series of test instruments has been built to provide a cost effective, yet powerful solution to hipot testing, whether on a production line or in a laboratory environment for compliance testing purposes. With the ability to perform either 5kV AC or 6kV DC dielectric strength tests and up to 500VA AC power output the Sentinel series can aid testing to many international safety standards including IEC, EN, UL and CSA to name but a few. With many features, the Sentinel series ensures testing is performed accurately, efficiently and safely.
Key Features
500 VA capability (Sentinel 500 series)
Earth bond / insulation resistance (Sentinel 501)
Sweep function for test analysis
Flexible / high resolution test parameters
Manual / automatic test modes
Remote and PLC control
Fast cut off and automatic discharge
Key Users
R&D Departments
Compliance Laboratories
Design Engineers
QA Departments
Production and Manufacturing Engineers
Technical Specifications
Sentinel 500/501 - AC Dielectric Strength
Programmable Voltage Range - 0.100 kV - 5.000 kV (2V/Step Resolution)
Frequency Independent of Supply - 50 or 60 Hz Selectable
Output Voltage Accuracy - ±1% + 5 Volts (No Load)
Voltage Display Resolution - 0.001kV
Current Measurement - 0.001mA - 100.0mA
Range and Accuracy - ±1.5% + 30 Counts when trip limit 1.11mA
Current Display Resolution - 0.001mA / 0.001mA / 0.1mA
Selectable Range of Pass/Fail Levels - 0.001mA - 100.00mA
Maximum Rated Current - 100mA (0.5kV ≤ V ≤ 5kV)
10mA (0.1kV ≤ V ≤ 0.5kV)
Arc Detection - yes