- Completely controlled by PC, can flexibly select the flame and graphite furnace atomizer (optional).
- Integrated floated optical platform design improves the optical system shock resistance stable.
- Eight light stands can be changed automatically and preheat the eight lights meantime as well as optimize the working condition of the hollow cathode lamp.
- Position adjusting: the flame burner can be set on the best position automatically.
- Fully automated wavelength scanning and peak searching.
- Complete safety chains protection equipment: warning and automatic safety protection for the wrong burner connection, leakage of the gas, under voltage of air and the abnormal flameout.
- Deuterium lamp and self-absorption background calibration.
- Measuring method: flame absorption method and emission method.
- Result printing: parameter, data result and diagram.
- Data processing: more than 500 data self-storage and cut-off storage, the analyzed result is stored in EXCEL format.
Technical Specifications
Technical Parameters
Wavelength Accuracy: ±0.15nm
Wavelength Receptivity: ≤0.04nm
Base Line Stability: ≤0.002A/30min (Cu)
Detection Limit: 0.004μg/ml (Cu)
Characteristic Concentration: 0.02μg/ml/1% (Cu)
Measurement Repeatability: ≤0.5%
Grating: 1800 lines/mm
Inflamer: All-metal titanium burner
Atomizer: Efficient glass atomizer
Lamp stand: 8
D2 background Calibration: When the background is 1 A, the background ability should be deducted not less than 50 times; self-absorption background deduction method