Product informationValidation data logger for thermal validation of processes with steam sterilizers (autoclaves). The logger features bendable metal probes and very thin temperature sensors (Ø 1.5 mm at the sensor tip) for fast response. Used with the AL 126 Endoscope Flow Control Adapter, it facilitates routine checks in RDG-E systems.
Benefits:- Bendable metal probe
- High measurement accuracy
- Real-time data transmission
- Fast response time
- Sterilizable data logger
Applications:- Validation measurements in steam sterilizers (autoclaves), washer-disinfectors and endoscope washers
- Routine checks in RDG-E when used with AL 126 adapter
Standards and directives:- ISO 17665 — Sterilization of health care products — Moist heat
- ISO 15883 — Washer-disinfectors
- 2014/53/EU — Radio Equipment Directive (RED)
- (EC) No 1935/2004 — Materials and articles intended to come into contact with food
Technical specifications:- Product number: 1340-6653
- Data memory: 2 x 33,440 measurement values
- Weight: approx. 120 g
- Accuracy: ±15 mbar (1 mbar ... 49 mbar); ±10 mbar (50 mbar … 150 mbar); ±15 mbar (151 mbar ... 2.049 mbar); ±10 mbar (2.050 mbar … 2.250 mbar); ±15 mbar (2.251 mbar ... 2.999 mbar); ±10 mbar (3.000 mbar … 3.250 mbar); ±15 mbar (3.251 mbar ... 3.499 mbar); ±20 mbar (3.500 mbar ... 4.000 mbar)
- Protection class: IP68
- Dimensions: ø 48 mm x 32 mm
- Resolution: 0.01 °C; 1 mbar
- Certificate / Factory calibration: 0 °C, +60 °C and +134 °C; 100 mbar and 3,100 mbar at +25 °C and 3,100 mbar at +134 °C
- Measured variables: Temperature, pressure
- Interval: 250 ms ... 24 h
- Battery: 3.6 V lithium battery, exchangeable
- Measuring range: Temperature: 0 °C … +150 °C; Pressure: 1 … 4,000 mbar
- Storage temperature: -20 °C ... +70 °C