MAGNETIC FIELD HiTESTER FT3470-51
Providing Robust Support for 3-Axis Magnetic Flux Density Measurement
Overview
Hioki magnetic field meters are designed to measure magnetic flux density and level of magnetic field exposure, and can be used to assess conformance to ICNIRP 2010 and IEC/EN62233 and in research on magnetic field exposure. The FT3470-51 measures magnetic flux density and magnetic field exposure generated by products manufactured at a customer site or produced by devices in a customer environment, and comes bundled with 1 sensor.
Key Features
• - Complies with ICNIRP 2010 guidelines as well as other relevant standards for evaluation testing.
• - Complies with IEC 62110/IEEE 644 as well as IEC 62233.
• - Bundled with 100 cm2 Sensor used with the IEC/EN 62233 standard analysis
• - User-selectable display units (T, A/m, and G)
• - Simple operation for easy measurement
• - Bundled with PC application software
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Magnetic flux density (Bandwidth) - 10 Hz to 400 kHz/ 10 Hz to 2 kHz/ 2 kHz to 400 kHz
Exposure level - General public/ Occupational
Display - Single axes X, Y, Z (2000 counts), Composite RMS value R (3464 counts), Magnetic flux density (unit: T, G, A/m), Exposure level (unit: %)
Magnetic flux density/Ranges and accuracy - [X, Y, Z axes] Effective measuring ranges: 2.000 μT to 2.000 mT, 4 ranges, Accuracy: ±3.5% rdg. ±0.5% f.s.
[R axis] Effective measuring ranges: 3.464 μT to 3.464 mT, 4 ranges, Accuracy: ±3.5% rdg. ±0.5% f.s.
[Valid measurement frequency range]
at 10 Hz-400 kHz mode: 50 Hz to 100 kHz
at 10 Hz-2 kHz mode: 50 Hz to 1 kHz