Product OverviewWinner319B is a bench-top spray laser particle size analyzer designed for droplet and aerosol size measurement. It uses the Fraunhofer diffraction principle with a parallel light path and a high-performance semiconductor laser (λ=635 nm). The split-type instrument (separate transmit and receive units) supports non-contact spray testing with a measurement range of 1–1000 µm (extendable to 0.1 µm) and is suitable for R&D laboratories, quality-control institutes and industrial test environments.
Main Specifications- Model Name: Winner319B
- Principle: Fraunhofer diffraction
- Instrument Structure: Split-type (separate transmit/receive ports)
- Data Acquisition Rate: ≥2 kHz
- Sampling Mode: Open style
- Power Supply: 220V, 50/60 Hz
- Dimensions: Transmit port L380 × W280 × H380 mm; Receive port L850 × W280 × H380 mm
- Net Weight: Transmit port 13 kg; Receive port 24 kg
Measurement Size- Measurement Range: 1–1000 µm (lower limit can be extended to 0.1 µm)
- Number of Channels: 78
- Accuracy (CRM D50): ≤1.0%
- Repeatability (CRM D50): ≤1.0%
- Reproducibility Between Instruments: <1.0% (CRM D50)
- Testing Area Length: 0.1–10 m (adjustable)
Optics- Light Source: Semiconductor laser λ=635 nm, power 1–40 mW (adjustable)
- Optical Bench Alignment: Automatic alignment system
- Lens Protection: Double gas airflow protection
- Laser Safety Class: 3B
Software Functions- Analysis Modes: free distribution, R-R distribution, log-normal distribution and category-specific statistical models
- Statistical Outputs: volume distribution and number distribution
- Statistical Comparison: multi-sample comparison for batch control and before/after process analysis
- Custom Analysis: user-defined parameters; calculate particle-size by percentage or percentage by particle-size; supports industry-specific characterization
- Report Export: Word, Excel, BMP, TXT, PDF
- Interface Languages: Chinese and English (additional languages on request)
- Optional: GMP compliance, angle testing
Advantages- Patented parallel optical testing and spectrum amplification with auxiliary photodetectors to collect scattering over wide angles and extend measurement range within compact space
- Airflow protection device to shield lenses and prevent contamination from droplets
- Split-type structure with adjustable test area for flexible, non-contact spray measurements
- Automatic optical-path alignment system for one-key correction of alignment shifts
- Multiple size-distribution models and conversion between volume and number distributions
Applications- Fire suppression spray testing
- Forest and agricultural pesticide spray evaluation
- Nebulizer and medical aerosol droplet testing
- Aircraft engine spray and aerosol research
- Nozzle development and spray prilling studies
- General aerosol and spray particle-size testing in enterprise, university and research institute laboratories
Technical Specifications- Model: Winner319B
- Measurement range: 1–1000 µm (extendable to 0.1 µm)
- Channels: 78
- Data acquisition rate: ≥2 kHz
- Accuracy / Repeatability (CRM D50): ≤1.0%
- Reproducibility between instruments: <1.0% (CRM D50)
- Testing area length: 0.1–10 m (adjustable)
- Light source: Semiconductor laser λ=635 nm, power 1–40 mW (adjustable)
- Optical alignment: Automatic alignment (3D optical-bench alignment)
- Lens protection: Double gas airflow device
- Laser safety class: 3B
- Instrument structure: Split-type (transmit & receive ports)
- Power supply: 220V, 50/60 Hz
- Report formats: Word, Excel, BMP, TXT, PDF
- Software: multiple distribution models, statistical comparison, custom analysis; Chinese & English interface