Product OverviewWinner3003 is a dry dispersion laser particle size analyzer designed for accurate and repeatable measurement of powders using Mie scattering theory. It integrates a converging Fourier optical path, a high-performance 639 nm laser and a high-sensitivity photodetector ring. The instrument uses air turbulent dispersion combined with a precision feeding device, a patented powder spray pump and an oil-free silent gas source to ensure complete dispersion of dry samples. It is particularly suited to powders that react or change shape in liquid media.
Key Features- Dry dispersion measurement based on Mie scattering.
- Converging Fourier (folded) optical path with high-performance laser λ = 639 nm, P > 2 mW and high-sensitivity detector ring.
- Turbulent dry dispersion with automatic sample introduction and SOP one-click operation.
- Patented powder spray pump and oil-free silent gas source for consistent dispersion.
- Automatic three-dimensional optical bench alignment and Class 1 laser safety.
Applications- Cement and ceramics production quality control.
- Pharmaceutical powders, pigments, dyes and fillers.
- Chemicals, catalysts and functional powders.
- Coal dust, additives, pesticides and energetic materials (as applicable to facility safety).
- Metal and non-metal powders, graphite, calcium carbonate, kaolin and other mineral powders.
- Materials sensitive to liquid media (e.g., magnetic powders, botanicals).
Advantages- Automated SOP workflow: automated dust removal, air supply, feeding, measurement and data processing to reduce operator variability.
- Robust dry dispersing system: turbulent patented dispersion and wear-resistant ceramic core for stable, long-life operation.
- Automatic optical alignment with precise hybrid stepper motor for micron-level positioning.
- Advanced optical design collecting scattering light across the measurement range for consistent accuracy.
- Flexible fitting models allowing free-distribution fitting and multiple statistical analyses.
Main Specifications- Model: Winner3003
- Principle: Mie scattering
- Measurement range: 0.1–300 μm
- Number of channels: 40
- Accuracy error: ≤1.0% (CRM D50)
- Repeatability error: ≤1.0% (CRM D50)
- Dispersion method: Turbulent dry dispersion, automatic sample introduction
- Test speed: fastest 10 seconds
- Sampling frequency: up to 10 kHz
- Light source: Laser λ = 639 nm, P > 2 mW
- Optical path: Reverse Fourier (folded)
- Optical bench alignment: Automatic (3D)
- Laser safety class: 1
- Dimensions: Main unit 765×450×325 mm; Dry module 260×245×295 mm
- Net weight: Main unit 26 kg; Dry module 7 kg
- Operation mode: SOP fully automated, one-click testing
Software & Output- Analysis models: free distribution, R-R, log-normal and other statistical fits.
- Batch-to-batch and process comparison tools for QC and cost control.
- Intelligent data processing to produce representative particle size results.
- Custom outputs: any D1–D100 characteristic sizes, cumulative and interval distributions.
- Outputs: particle size metrics, differential and cumulative distributions as data and graphs (histograms/curves).
- Report export: Word, Excel, BMP, TXT, PDF.
- One-click automation of dust collection, air supply, feeding, measurement and processing.
- Interface languages: English and Chinese (additional languages configurable).
- GMP compliance: optional.
Technical Specifications- Model: Winner3003
- Measurement range: 0.1–300 μm
- Principle: Mie scattering
- Light source: Laser λ = 639 nm, P > 2 mW
- Channels: 40
- Accuracy error: ≤1.0% (CRM D50)
- Repeatability: ≤1.0% (CRM D50)
- Sampling frequency: up to 10 kHz
- Test speed: ≥10 s (fastest)
- Dispersion: Turbulent dry dispersion, automatic feed
- Optical path: Reverse Fourier (folded)
- Optical bench: Automatic alignment
- Laser safety class: 1
- Dimensions: Main 765×450×325 mm; Dry module 260×245×295 mm
- Net weight: Main 26 kg; Dry module 7 kg
- Output formats: Word, Excel, BMP, TXT, PDF