OverviewThe CellaCombustion PA 47 is a non-contact two-colour pyrometer designed for temperature measurement of sooty flames in coal-fired power stations and incineration plants. The instrument detects near-infrared radiation at two wavelengths and applies an algorithm to correct wavelength-dependent soot emission and optical flame density, providing reliable flame temperature readings for combustion control, emission reduction and slagging minimization.
Special features- Measuring range: 700 – 1700 °C
- Focusable precision lenses for accurate measurement distance adjustment
- Broadband anti-reflective optics
- 4-digit red LED display, digit height 8 mm, readable at distance
- Test current output for diagnostics
- Standard interfaces: digital I/O, two analogue outputs, USB and RS-485
Properties / Options (summary)Measuring range: 700 - 1700 °C
Focus distance: 0.4 m - ∞ (standard PZ 20.01)
Spectral range: 0.8 / 1.05 µm
Distance ratio: 80 : 1
Sighting: through-the-lens, laser pilot light or built-in video camera option
Example version: CellaCombustion PA 47 AF 1
Technical data (measurement parameters)Sensor: double photodiode (two-colour)
Response time t98: ≤ 20 ms (T > 650 °C); ≤ 10 ms (T > 750 °C)
Measurement uncertainty: 1 % of temperature reading
Repeatability: 2 K
Resolution (interface): 0.1 K; display resolution: 1 K
Optical variantsStandard objective PZ 20.01: focus 0.4 m - ∞
Distance ratio: 80 : 1
Measuring field example at 400 mm: Ø 5.00 mm
Electrical parametersAnalogue outputs: 2 × 0(4)–20 mA, linear, switchable, scalable
Resistance: ≤ 500 Ω
Analog input: 0–10 V
Switching outputs: 2 × open collector 24 V DC, ≤ 30 mA
Switching inputs: 2 × up to 24 V
Interfaces: USB, RS-485
Power supply: 24 V DC (+10 % / -20 %); current ≤ 175 mA
Camera / SightingVideo output (camera version): composite PAL, 1 Vpp, 75 Ω
Sighting options: through-the-lens, laser pilot light, built-in camera (video)
Features & Displays4-digit LED display (red), status LEDs (2), 3 control buttons
On-board functions: contamination monitoring threshold, soot factor, ATD (automatic temperature detection), measured-value/current simulation, overtemperature signaling (analogue output behaviour if internal temp > 80 °C)
General parametersConnection: M12, 8-pin
Ambient temperature: 0 – 65 °C
Storage temperature: -20 – +80 °C
Humidity: up to 95 % r.H. (non-condensing)
Dimensions: Ø 65 mm × 220 mm (max. 277 mm)
Housing: stainless steel V2A (1.4305)
Protection class: IP65
Weight: approx. 0.9 kg
Adjustable parameters (examples)Measuring range, measured variable and scaling, smoothing filter, memory mode, emissivity, transmittance, reflected ambient radiation compensation, alarm limits and modes, ATD parameters (optional), linearization, LED settings, simulated current/temperature, unit °C/°F, measurement channel selection, contamination threshold, soot factor. Camera models: TBC, white balance.
Scope of delivery (typical)Pyrometer CellaCombustion PA (PA 47)
Connection cable VK 02/A (5 m)
Video cable VK 02/F (5 m) for camera models
USB cable VK 11/D (1.8 m)
Software CellaView
Short manual
SoftwareCellaView for configuration, data readout and diagnostics
Technical specifications- Model: CellaCombustion PA 47
- Measuring range: 700 – 1700 °C
- Spectral range: 0.8 / 1.05 µm
- Distance ratio: 80 : 1; focus 0.4 m – ∞; standard objective: PZ 20.01
- Response time: ≤ 20 ms (T > 650 °C); ≤ 10 ms (T > 750 °C)
- Accuracy: 1 % of reading; repeatability: 2 K
- Outputs / Interfaces: 2 analogue outputs (0/4–20 mA), digital I/O, USB, RS-485
- Power: 24 V DC (+10 % / -20 %); current ≤ 175 mA
- Camera option: composite PAL video output
- Protection / Housing: IP65, stainless steel V2A
- Dimensions / Weight: Ø 65 mm × 220 mm (max. 277 mm) / approx. 0.9 kg