Class C pyranometer SMP3-A
analogdigitalwith amplified output

Class C pyranometer - SMP3-A - Kipp & Zonen - analog / digital / with amplified output
Class C pyranometer - SMP3-A - Kipp & Zonen - analog / digital / with amplified output
Class C pyranometer - SMP3-A - Kipp & Zonen - analog / digital / with amplified output - image - 2
Class C pyranometer - SMP3-A - Kipp & Zonen - analog / digital / with amplified output - image - 3
Class C pyranometer - SMP3-A - Kipp & Zonen - analog / digital / with amplified output - image - 4
Class C pyranometer - SMP3-A - Kipp & Zonen - analog / digital / with amplified output - image - 5
Class C pyranometer - SMP3-A - Kipp & Zonen - analog / digital / with amplified output - image - 6
Add to favorites
Compare this product

Characteristics

IEC 61724-1
class C
Output signal
analog, RS-485, digital, current output, 4-20 mA, Modbus RTU, with amplified output
Options
ISO 9060, spectrally flat, digital, smart

Description

Overview
The Kipp & Zonen SMP3-A is a smart pyranometer for industrial and meteorological applications, providing amplified analogue and digital outputs with low maintenance requirements. Based on CMP3 technology, the SMP3 adds a Modbus® interface, improved response time and temperature-corrected measurements. Its wide supply range (5–30 VDC) and protection against over-voltage, reverse polarity and short circuits simplify integration into measurement stations. Identical sensitivity and connections across units enable quick replacement during recalibration.

Benefits
  • ISO / IEC classification: Spectrally Flat Class C (ISO 9060) with ISO / IEC 17025 calibration.
  • Low maintenance: internal drying cartridge — no replacement needed for 10 years; 5-year warranty and high MTBF.
  • Smart outputs and low power: Modbus RTU (2-wire RS-485) plus analogue outputs (4–20 mA), temperature correction and minimal post-recalibration adjustments.


Technical specifications
  • Non‑stability: < ±1% (change/year)
  • Drying cartridge & maintenance: internal, no replacement necessary
  • Zero offset B: < ±5 W/m²
  • Operating humidity: 0 to 100 %
  • Irradiance saturation: 2000 W/m² (max)
  • Enclosure material: aluminium, anodised
  • Digital outputs: Modbus RTU 2‑wire RS‑485
  • Weight: 300 g (0.7 lb)
  • Non-linearity: < ±3% (0 to 100 W/m²)
  • Directional response: < 20 W/m² (up to 80° with 1000 W/m² beam)
  • Zero offset A: < ±15 W/m²
  • Spectral range / accuracy: 300 to 2800 nm
  • Ingress protection: IP67
  • Classification: Spectrally Flat Class C (ISO 9060:2018)
  • Temperature correction: < ±3% (-20 to +50 °C)
  • Operating temperature range: -40 to +70 °C
  • Response time: < 1.5 s (63%), < 12 s (95%)
  • Analogue outputs: 0 to 1 V (V-model) or 4 to 20 mA (A-model)
  • Cable lengths available: 10, 25, 50, 100 m
  • Sensitivity: N.A.

Catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.