Product OverviewSmart cartridge designed by Kunak that integrates a sensor capable of accurately and reliably measuring Carbon monoxide levels.
- GasPlug TECHNOLOGY | PATENTED DESIGN
Where is it found?CO is found in fumes produced any time fuel burns in cars or trucks, small engines, stoves, lanterns, grills, fireplaces, gas ranges, or furnaces. Improperly ventilated appliances and engines, particularly in a tightly sealed or enclosed space, may allow carbon monoxide to accumulate until dangerous levels.
Why is it harmful?Carbon monoxide or CO is an odourless and colorless pollutant gas that can cause death. When CO is inhaled into the body, it mixes with the blood avoiding the oxygen absorption. The CO exposure of a person over a certain period of time can cause illness and even death.
CO CartridgeThe carbon monoxide cartridge has a built-in electrochemical sensor with very low noise electronics allowing the measurement from very low concentrations (ppb) up to several ppm. There are two measurement ranges to cover different applications:
- Type A measuring most common concentrations range to 12,000 ppb.
- Type B, a higher range version that can measure up to 500 ppm, with reduced accuracy at low concentrations.
This cartridge is very stable over time and, in non-extreme environments, can last several months longer than specified.
Technical specificationsType | Electrochemical
Unit of measurement | µg/m³, ppb (A)
mg/m³, ppm (B)
Measurement range (1) | 0 - 12,000 ppb (A)
0 - 500 ppm (B)
Resolution (2) | 1 ppb (A)
0.01 ppm (B)
Operating temperature range (3) | -30 to 50 ºC
Operating RH range (4) | 0 to 99 %RH
Recommended RH range (4) | 15 to 90 %RH
Operating life (5) | > 24 months
Guarantee range (6) | 1,000 ppm
Limit of Detection (LOD) (7) | 10 ppb (A)
0.02 ppm (B)
Repeatability (8) | 20 ppb (A)
0.05 ppm (B)
Response time (9) | < 30 sec (A)
< 180 sec (B)
Typical accuracy (11)(12) | ± 80 ppb (A)
± 0.1 ppm (B)
Typical precision R² (10) | > 0.85
Typical slope (10) | 0.78 - 1.29
Typical intercept (a) (10) | -50 ppb ≤ a ≤ +50 ppb (A)
-0.1 ppm ≤ a ≤ +0.1 ppm (B)
DQO - Typical U(exp) (13) | < 20%
Typical Intra-model variability (14) | < 3 ppb (A)
< 0.05 ppm (B)
Characteristics / Technical Specifications- Sensor type: Electrochemical
- Measurement ranges: 0 - 12,000 ppb (Type A), 0 - 500 ppm (Type B)
- Resolution: 1 ppb (A), 0.01 ppm (B)
- Operating temperature: -30 to 50 ºC
- Operating humidity: 0 to 99 %RH (recommended 15 to 90 %RH)
- Operating life: > 24 months
- Guarantee range: 1,000 ppm
- Limit of Detection: 10 ppb (A), 0.02 ppm (B)
- Repeatability: 20 ppb (A), 0.05 ppm (B)
- Response time: < 30 sec (A), < 180 sec (B)
- Typical accuracy: ± 80 ppb (A), ± 0.1 ppm (B)
- Typical precision R²: > 0.85
- Typical slope: 0.78 - 1.29
- Typical intercept: -50 ppb ≤ a ≤ +50 ppb (A), -0.1 ppm ≤ a ≤ +0.1 ppm (B)
- DQO - Typical U(exp): < 20%
- Typical intra-model variability: < 3 ppb (A), < 0.05 ppm (B)