Product overviewThe Memmert ICO180 provides a controlled CO2 incubation environment for reproducible cell and tissue culture results. Precise CO2 and humidity control, multi-zone temperature conditioning and advanced connectivity enable secure processes and traceable parameter management.
Range of functionsKey operating and monitoring features include:
- Calibration of temperature, humidity and CO2 with three freely selectable setpoints
- Interfaces: USB-C and Ethernet LAN for data transfer and network integration
- Flexible timers: immediate start, SetpointWAIT (start after reaching setpoints), and event-based start
- Extensive language options and configurable appliance settings (date/time, display brightness, DistanceView, network and alarm settings)
O2 (only with T6 option)Anaerobic oxygen control via filtered nitrogen inlet. Zirconium-oxide O2 sensor with integrated monitoring and error reporting. Setting range 1–20 % with 0.1 % adjustment precision.
TemperatureLarge-area multi-zone heating with additional door, seal and rear-wall heating ensures uniform conditions. Setpoint range +20 to +50 °C; usable down to 5 °C above ambient. Built-in sterilisation programme: 60 minutes at 180 °C (sensors may remain in place). Display resolution 0.1 °C; spatial variation at 37 °C ±0.3 K; temporal fluctuation ±0.1 K.
HumidityStandard passive humidification via integrated water tank with condensate return; humidity limiting by Peltier element. Option K7 provides active closed-loop humidity control with capacitive sensor, micro-dosing pump for hot-steam humidification and diaphragm pump dehumidification. Accuracy 0.5 % RH; setting range 40–98 % RH and rh-Off.
CO2Dual-beam NDIR CO2 control with pressure-compensated measurement and integrated monitoring. Filtered CO2 supply for process reliability. Temporal fluctuation ±0.2 %; setting range 0–20 %.
SafetyPower-failure behaviour configurable per mode (manual, program). Dual Pt1000 temperature sensors (DIN Class A) for mutual monitoring and automatic takeover. Visual and acoustic alarms.
Standard equipmentDelivered with door-open detection (V5) and factory calibration options (Standard I, K7 II, T6 III). Includes CO2 connection set, water tank (B38737) and stored sterilisation programme. Fully insulated stainless-steel door with 2-point compression lock and inner glass door with sampling port.
Stainless steel interiorInternal dimensions (W x H x D): 600 x 590 x 500 mm; inner cover 37.5 mm; usable volume 177 L; max internals 6; chamber load max 90 kg; shelf load max 15 kg.
Textured stainless steel casingExternal dimensions (W x H x D): 748 x 793 x 689 mm. Appliance on 4 adjustable feet.
Electrical data230 V, 50/60 Hz — approx. 1500 W; 115 V, 50/60 Hz — approx. 1500 W.
Ambient conditionsMinimum clearances: rear 15 cm, top 20 cm, sides 5 cm. Ambient temperature 10 °C to 35 °C. Installation altitude up to 2,000 m. Overvoltage category II; pollution degree 2.
Packing / shipping dataTransport upright. Customs tariff number: 8419 8998. Country of origin: Federal Republic of Germany. WEEE-Reg.-No.: DE 66812464. Packed dimensions approx. 925 x 1072 x 854 mm. Net weight approx. 98 kg; gross weight approx. 116 kg.
CertificationsStandard units are safety-approved and bear CE marking.
Technical specifications- Brand: Memmert
- Model: ICO180
- Usable volume: 177 L
- Internal dimensions (W x H x D): 600 x 590 x 500 mm
- External dimensions (W x H x D): 748 x 793 x 689 mm
- Temperature setpoint range: +20 to +50 °C (usable from +5 °C above ambient to +50 °C)
- Temperature resolution: 0.1 °C; spatial homogeneity at 37 °C: ±0.3 K; temporal fluctuation ±0.1 K
- CO2 range: 0–20 % (±0.2 % temporal fluctuation)
- Humidity: 40–98 % RH (rh-Off); accuracy 0.5 % RH; active control option K7
- O2 (option T6): 1–20 %; accuracy 0.1 %
- Control/display: 6.86" IPS multi-touch (480×1280)
- Max shelves: 6; max chamber load: 90 kg; max per shelf: 15 kg
- Power: approx. 1500 W (230 V / 115 V)
- Net/Gross weight: approx. 98 kg / 116 kg