Short descriptionThe perfect environment for superior cell growth. Secure, reproducible CO2 incubation with precise humidity control, intelligent hygiene design and advanced connectivity features.
Range of functionsCalibrate: Three freely selectable values: temperature, humidity, and CO2
Interfaces: USB-C; Ethernet LAN
Timer: Timer - Process time starts immediately when the programme starts; Timer SetpointWAIT - Process time only starts after all setpoint values have been reached; Timer on Event - Process time starts at a defined point in time
Language settings:- English (default)
- German
- French
- Spanish
- Portuguese
- Chinese
- Arabic
- Hindi
- Polish
- Czech
- Hungarian
- Russian
- Italian
- Serbian
Appliance functions: Date and time (time zones, summer/winter time); Display brightness; DistanceView (1–4 freely selectable parameters, activation time adjustable); Network settings (IP, subnet, DHCP); Language and volume settings (buzzer/key tone); Power failure behaviour; Timer functions
Adjustable parameters: Temperature; CO2; Humidity (requires option K7); O2 (requires option T6)
Control unit: 6.86" IPS touch display (480 × 1280 pixels) with multi-touch function for graphical display of actual/setpoint values and parameter curves.
O2 (only with T6 option)Control system: Anaerobic O2 control for targeted oxygen reduction via nitrogen inlet. Zirconium oxide sensor with integrated monitoring and error display. Filtered N2 supply for safe process conditions.
Adjustment precision: 0,1 %
O2 - Setting range: 1 to 20 %
TemperatureTemperature control system: Large-area multi-zone multifunctional all-round heating with additional door, seal and rear wall heating
Setting temperature range: +20 to +50 °C
Working-temperature range: 5 °C above ambient temperature up to +50 °C. Standard sterilisation programme: 60 minutes at 180 °C (without removing the sensors).
Setting accuracy temperature: 0.1 °C
Temperature variation in chamber: at + 37 °C +/- 0.3 K
Temperature fluctuations with time: at 37 °C +/- 0.1 K
HumidityHumidification system: Passive humidification by integrated water tank with condensate return via magnetically guided sealing lip. Humidity limitation by Peltier element. / Active closed-loop humidity control (option K7) with capacitive sensor for precise measurement and complete documentation. Humidification via self-priming micro-dosing pump with hot steam generation, dehumidification via powerful diaphragm pump.
Accuracy: 0,5% rh
Setting range: 40 to 98 and rh-Off
CO2Control system: Precise CO2 control with dual-beam NDIR technology, integrated monitoring and pressure-compensated measurement. Filtered CO2 supply for high process reliability.
Temporal fluctuation: ± 0,2 %
Setting range: 0 to 20 %
SafetyPower failure: Manual mode / Power failure less than 24 hours: Active process is terminated, parameters remain stored. Power failure longer than 24 hours: Operation continues with previously set parameters. Program mode (power failure >1 hour) / Failure less than 1 hour: Programme continues with remaining runtime. Failure longer than 1 hour: Programme is stopped; query for continuation required.
Temperature monitoring: Dual Pt1000 sensor, DIN Class A, for mutual monitoring and automatic takeover in the event of a fault
Alarm: visual and acoustic
Standard equipmentStandard option: Door-open detection | V5
Factory calibration certificate III (Option T6): +37 °C, 5 %CO2, 90 %RH, 10 %O2
Factory calibration certificate II (Option K7): +37 °C, 5 %CO2, 90 RH
Factory calibration certificate I (Standard): +37 °C, 5 %CO2
Additional accessories: CO2 connection set, hose with connection coupling and hose clamp; Water tank | B38737; Water tank including connection hose; Stored sterilisation programme: 60 minutes at +180 °C (without removing the sensors)
Door: fully insulated stainless steel door with 2-point locking (compression door lock)
Inner door: inner glass door with opening (8 mm Ø) to take gas sample
Internals: 1 perforated stainless steel shelf/shelves
Stainless steel interiorDimensions (w(A) x h(B) x d(C)): 560 x 480 x 400 mm
Inner cover: 37,5 mm
Volume: 108 l
Max. number of internals: 5
Max. loading of chamber: 75 kg
Max. loading per internal: 15 kg
Textured stainless steel casingDimensions (w(D) x h(E) x d(F)): 708 x 683 x 589 mm
Installation: 4 adjustable appliance feet
Electrical data230 V, 50/60 Hz - approx. 1300 W
115 V, 50/60 Hz - approx. 1300 W
Ambient conditionsSet Up: The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature: 10 °C to 35 °C
Altitude of installation: max. 2,000 m above sea level
Overvoltage category: II
Pollution degree: 2
Packing / shipping dataTransport information: The appliances must be transported upright
Customs tariff number: 8419 8998
Country of origin: Federal Republic of Germany
WEEE-Reg.-No.: DE 66812464
Dimensions approx incl. carton (w x h x d): 825 x 967 x 815 mm
Net weight: approx. 70 kg
Gross weight carton: approx. 85 kg
Caractéristiques / spécifications techniques- Brand: Memmert
- Model: ICO105
- Volume: 108 l
- Internal dimensions (w x h x d): 560 x 480 x 400 mm
- External dimensions (w x h x d): 708 x 683 x 589 mm
- Setting temperature range: +20 to +50 °C
- Temperature accuracy: 0.1 °C; variation at 37 °C: ±0.3 K; fluctuation with time at 37 °C: ±0.1 K
- Humidity setting range: 40 to 98 % RH (and rh-Off); humidity accuracy: 0.5 % RH
- CO2 setting range: 0 to 20 %; temporal fluctuation ±0.2 %
- O2 option (T6): 1 to 20 %, adjustment precision 0.1 %
- Display: 6.86" IPS touch (480 × 1280 px) multi-touch
- Power: approx. 1300 W (230 V or 115 V variants)
- Net weight: approx. 70 kg; Gross weight carton approx. 85 kg
- Max. loading chamber: 75 kg; max internals: 5; max loading per internal: 15 kg
- Certification: CE
- Country of origin: Federal Republic of Germany