PSPT-03 Catheter Friction Performance Tester is specialized in testing the fastness of the hydrophilic coating, surface sliding performance/lubricity, coating friction and coefficient of friction of intravascular catheters, non-vascular catheters, guidewires, etc., such as urinary catheters, drainage tubes, tracheal intubation tubes, balloon catheters, etc.
Standards
T-CAMDI 021-2019 Single-use hydrophilic coated guidewire
YY/T 1898-2024 Intravascular catheter and guidewire - Testing method for fastness of the hydrophilic coating
YY/T 1536-2017 Catheters other than intravascular catheters-standard test model for evaluating the surface sliding performance
Unique cyclic friction test mode, and supports multiple one-way up and multiple one-way down test modes
Support friction times setting to meet multiple friction tests
Static friction, kinetic friction, static coefficient of friction , kinetic coefficient of friction can be tested for one time
Automatic clamping device, the force is settable freely, to ensure the stability of the test
Adopt circulating water bath temperature control design to realize the test under different temperature conditions.
Silicone pads of standardized hardness to provide a test environment close to human tissue.
Small range sensor to ensure the accuracy of test data
The equipment supports one-key automatic water supply and drainage function, which ensures a clean water bath environment
0.5%FS high-precision testing, effectively ensures the accuracy and repeatability of test results
Equipped with advanced displacement control system, displacement accuracy can up to 0.1mm