Gas chromatograph DHA
laboratoryfor chemical applicationstemperature-programmable

Gas chromatograph - DHA - Shimadzu France - laboratory / for chemical applications / temperature-programmable
Gas chromatograph - DHA - Shimadzu France - laboratory / for chemical applications / temperature-programmable
Add to favorites
Compare this product

Characteristics

Type
gas
Applications
laboratory, for chemical applications
Other characteristics
analytical, capillary, with column oven, temperature-programmable

Description

Overview
The purpose of detailed hydrocarbon analysis (DHA) is to determine the bulk hydrocarbon group-type composition (PONA: Paraffins, Olefins, Naphthenes and Aromatics) of gasoline and other spark-ignition engine fuels containing oxygenates (methanol, ethanol, MTBE, ETBE, TAME) according to ASTM-D6730. PONA identification is performed by matching retention indices with n-alkane standards. Chromatograms contain a large number of peaks; accurate and reproducible retention indices are essential to avoid erroneous PONA ratios. The Nexis GC-2030 with Advanced Flow Controller enables precise flow control and reproducible retention indices. Shimadzu offers DHA services and methods conforming to ASTM-D5134, ASTM-D6729, ASTM-D6730 and ASTM-D6733.

Analytical Conditions
Equipment: PONA System (Nexis GC-2030AF)
Column: Rtx-PONA, 100 m x 0.25 mm x 0.5 μm
Pre-column: Pre-PONA, 2 m x 0.25 mm x 1 μm
Column temperature program: 5 °C (hold 10 min) → 5 °C/min → 48 °C (hold 49 min) → 1.2 °C/min → 200 °C (5 min)
Carrier gas: He; split 1:150 (initial 5 min) then 1:20 for remainder
Sample size: 0.2 μL
Injector temperature: 250 °C
Detector temperature: 250 °C
Cryogenic option: CRG-2030 with liquid CO2

Example of a DHA analysis chromatogram (summary data)
Group Type | Total Mass% | Total Vol.% | Total Mol%
Paraffins | 15.868 | 17.1143 | 15.9838
Iso-Paraffins | 19.495 | 20.8586 | 19.1926
Olefins | 17.3556 | 18.5952 | 19.8748
Naphthenes | 20.6483 | 20.0522 | 21.3048
Aromatics | 23.5425 | 20.3256 | 21.8861
Total C14+ | 3.0907 | 3.054 | 1.7579
Total Unknowns | 0 | 0 | 0
Total Oxygenates | 0 | 0 | 0
Sample Total | 100 | 100 | 100
Multi-substituted Aromatics: 13.7740 (mass %), 11.8683 (vol. %)
Average Molecular Weight: 112.8396
Reid Vapor Pressure @ 100°F: 1.9378
Calculated Octane Number: 77.0250

Boiling Point Summary (by mass and vol.)
Property | IBP | T10 | T50 | T90 | FBP
BP by mass (°F) | 82.11 | 153.14 | 227.16 | 402.10 | 488.66
BP by vol. (°F) | 82.11 | 146.25 | 274.10 | 401.72 | 488.66
Percent Carbon: 86.7143 | Percent Hydrogen: 13.2857 | Bromine Number (Calc): 22.8036

Example of summary by group type & carbon number (mass %)
Carbon | Paraffins | I-Paraffins | Olefins | Naphthenes | Aromatics | Unknowns | Total
C1 | 0 | 0 | 0 | 0 | 0 | 0 | 0
C2 | 0 | 0 | 0 | 0 | 0 | 0 | 0
C3 | 0 | 0 | 0 | 0 | 0 | 0 | 0
C4 | 0 | 0 | 0 | 0 | 0 | 0 | 0
C5 | 2.61412 | 0.39222 | 2.12534 | 1.02294 | 0 | 0 | 6.15463
C6 | 1.75622 | 1.74186 | 3.4275 | 1.83233 | 1.5174 | 0 | 10.27531
C7 | 1.78205 | 3.41605 | 4.85244 | 3.76943 | 1.02589 | 0 | 14.84587
C8 | 1.68547 | 4.74174 | 2.49244 | 6.10152 | 3.72353 | 0 | 18.74469
C9 | 1.74756 | 5.41724 | 3.0002 | 5.20934 | 5.63764 | 0 | 21.01198
C10 | 1.60532 | 3.77676 | 1.45765 | 1.93379 | 6.87431 | 0 | 15.64782
C11 | 1.56754 | 0.0091 | 0 | 0.01491 | 1.45758 | 0 | 3.04912
C12 | 1.56411 | 0 | 0 | 0.76403 | 3.3061 | 0 | 5.63424
C13 | 1.54563 | 0 | 0 | 0 | 0 | 0 | 1.54563
Total C14+ 3.09070, Oxygenates 0.00000, Total Unknowns 0.00000, Grand Total 100.00000

Available software / downloads
Software for PONA/DHA analysis: PONAsolution for LabSolutions (brochure and documentation available).

Technical characteristics / specifications
  • Application: Detailed Hydrocarbon Analysis (DHA) / PONA composition according to ASTM standards (ASTM-D5134, D6729, D6730, D6733)
  • Recommended system: PONA System (Nexis GC-2030AF)
  • Column: Rtx-PONA, 100 m x 0.25 mm x 0.5 μm
  • Pre-column: Pre-PONA, 2 m x 0.25 mm x 1 μm
  • Column temperature program: 5 °C (hold 10 min) → 5 °C/min → 48 °C (hold 49 min) → 1.2 °C/min → 200 °C (5 min)
  • Carrier gas: Helium; split ratios: 1:150 (initial 5 min) then 1:20
  • Sample volume: 0.2 μL
  • Injector temperature: 250 °C
  • Detector temperature: 250 °C
  • Cryogenic option: CRG-2030 with liquid CO2
  • Output: Group-type summary (mass%, vol%, mol%), carbon-number distribution, boiling point and derived properties (Average MW, RVP, Calculated Octane)
  • Software: PONAsolution for LabSolutions (dedicated PONA/DHA analysis software)
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.