OverviewMEPS (Micro Extraction by Packed Sorbent) miniaturizes conventional SPE packed-bed devices from millilitre to microlitre bed volumes. MEPS supports reversed-phase, normal-phase, mixed-mode and ion-exchange chemistries and is available in common SPE phases.
MEPS Barrel Insert and NeedleThe Barrel Insert and Needle (BIN) contains the stationary phase and is integrated into the syringe needle, enabling direct extraction within the syringe assembly.
Sample size and sensitivityMinimum sample volumes can be 10 µL. Larger sample volumes can be concentrated by taking multiple aliquots (for example 100 µL or 250 µL aliquots to accumulate ≥1 mL).
AutomationMEPS can be automated (AOC-MEPS) for on-line extraction and injection with a single device, reducing processing time and operator interventions in routine workflows.
Sorbent lifeBIN lifetime depends on sample matrix. Example: C18 analysis of whole plasma typically allows ~25–100 samples before BIN replacement; cleaner matrices give significantly longer life.
CarryoverThe small quantity of phase in the BIN is easily washed between samples to minimise carryover; automation permits washing during previous-sample runtime.
Flexible and easy to useThe sorbent bed dimensions deliver performance comparable to conventional SPE for similar sample types. The AOC-MEPS workflow integrates sample preparation and analysis on a single platform.
AOC-MEPS system — injection volumesTypical automated injections are 50–200 µL of elution solvent into the analytical instrument. Use a large-volume injection technique that removes solvent while condensing target analytes.
PTV Inlet Liner for AOC-MEPS System (AOC-20i + GC-2010 + PTV configuration)OD [mm] | ID [mm] | Length [mm] | P/N
3.5 | 2.5 | 95 | 221-74830-09
AOC-20i/20s — MEPS Syringe for AOC-20i/20sSyringe Volume | PTFE Tipped Plunger | P/N | QTY | Replacement Plunger P/N
100 µl | yes | 221-74830-01 | 1 | 221-74830-10
Available phases (AOC-20i/20s)C18 | 5 | 221-74830-03
Silica | 5 | 221-74830-04
C8 + SCX | 5 | 221-74830-05
C2 | 5 | 221-74830-06
C8 | 5 | 221-74830-07
PDVB | 5 | 221-74830-02
SDVB | 5 | 221-74830-11
HDVB | 5 | 221-74830-12
Development Kit - contains 1 each of C18, C8, Silica, C8+SCX, C2 | 1 of each phase | 221-74830-08
AOC-5000 — MEPS Syringe for AOC-5000Syringe Volume | PTFE Tipped Plunger | P/N | QTY | Replacement Plunger P/N
100 µl | yes | 221-75202-01 | 1 | 221-75202-03
250 µl | yes | 221-75202-02 | 1 | 221-75202-04
Available phases (AOC-5000)C18 | 5 | 221-75198-01 | 221-75199-01
Silica | 5 | 221-75198-02 | 221-75199-02
C8 + SCX | 5 | 221-75198-03 | 221-75199-03
C2 | 5 | 221-75198-04 | 221-75199-04
C8 | 5 | 221-75198-06 | 221-75199-06
Development Kit - contains 1 each of C18, C8, Silica, C8+SCX, C2 | 1 of each phase | 221-75198-05 | 221-75199-05
Technical specifications- Product family: MEPS (Micro Extraction by Packed Sorbent)
- Suitable chemistries: reversed-phase, normal-phase, mixed-mode, ion exchange
- Typical minimum sample volume: 10 µL (concentration via multiple aliquots possible)
- Typical automated injection volumes (AOC-MEPS): 50–200 µL elution solvent
- Sorbent/BIN life example: C18 with whole plasma ~25–100 samples (matrix dependent)
- PTV inlet liner (AOC-MEPS): OD 3.5 mm, ID 2.5 mm, Length 95 mm, P/N 221-74830-09
- AOC-20i/20s syringe: 100 µL (P/N 221-74830-01); replacement plunger P/N 221-74830-10; available phases and P/Ns listed above
- AOC-5000 syringes: 100 µL (P/N 221-75202-01) and 250 µL (P/N 221-75202-02); replacement plungers and phase P/Ns listed above
- Pack sizes: most MEPS phase packs contain 5 units